Operation
TIME
RPM
RCF
[TIME] key
■
Runtime, parameter t/hms.
h: hours. Can be set from 1 h to 99 h, in 1-hour increments.
m: minutes. Can be set from 1 min to 59 min, in 1-minute increments.
s: seconds. Can be set from 1 s to 59 s, in 1-second increments.
■
Continuous run
‘
@
’
■
Set the beginning of runtime counting. This setting is only possible when
the
‘Dual time
mode’ function is activated
Chapter 7.3.5.12 ‘Dual time mode’
on page 66. This function is deactivated ex works.
It can be set whether the runtime should begin immediately after the
start of the centrifugation run or only start counting after reaching the set
speed.
‘Timing begins at Start’: The runtime begins counting immediately after
the centrifugation run begins.
‘Timing begins at Speed’: The runtime only begins counting after
reaching the set speed.
‘Timing begins at Speed’: This is indicated on the display by the
symbol , to the left next to the time.
[RPM] key
Speed, parameter RPM.
Can be set from 50 RPM to the maximum rotor speed (Nmax), in incre-
ments of 10. Maximum rotor speed
Appendix A ‘Rotors and accessories’
[RCF] key
■
Relative centrifugal force, parameter RCF.
The RCF is displayed in brackets
>
<
. The LED in the button lights up. A
number value can be set which results in a speed between 50 RPM and
the maximum rotor speed (Nmax). Adjustable in increments of 1.
■
Centrifuging radius, parameter RAD. Can be set from 10 mm to 330
mm, in 1-mm increments. Centrifuging radius
accessories’ on page 86. The centrifuging radius is not automatically
taken from the rotor code / rotor information, but must be entered man-
ually.
■
Query of the integral RCF. It is only possible to query the integral RCF if
the display of the integral RCF is activated