Safety information
Danger posed by selection of
unsuitable parameters for dual
centrifugation
Damage due to incorrect
loading for dual centrifugation
WARNING
Danger posed by selection of unsuitable parameters for dual
centrifugation!
The maximum permissible speed (RPM) for ZentriMix appli-
cations depends on the sample composition and sample
container. Every ZentriMix sample container can only be
used up to a certain rotational speed depending on the
design and application; this must not be exceeded under
any circumstances! Otherwise, there is a danger that the
sample container will not withstand the stress, and the
device will be damaged due to the resulting unbalance; the
heat that arises cannot be compensated by the cooling
system, resulting in the sample container becoming very hot.
Only remove such a container after a sufficient cooling-off
time while wearing appropriate protective clothing.
−
ZentriMix applications are to be carried out according to
the protocol authorized by the manufacturer.
−
Every application not yet described by a manufacturer
protocol must be checked and validated for any hazard
potential before its use.
−
The correct parameter input must always be checked.
−
Do not leave the device unsupervised.
WARNING
Damage due to incorrect loading for dual centrifugation!
The two rotor receptacles must be loaded strictly symmetri-
cally with the same weight.
−
Observe the loading rules.
rotor for dual centrifugation’ on page 33!
−
Observe the device as it starts up. If unusually high
unbalance occurs, stop the run immediately by pressing
the
[Stop/Open] key twice and check the load. If you
cannot determine that the load is incorrect, put the
device out of operation and inform the Service depart-
ment.