Operation
Fig. 20: Removing the rotor
5.
Carefully remove the rotor upward and out of the device.
🡆
The interior (centrifuge chamber) is now accessible.
NOTICE
If the rotor is incorrectly installed or fastened, it could end up
coming loose.
When installing the rotor, make sure that the driver of the
rotor shaft is sitting correctly in the rotor groove.
On the top side of the rotor, there are markings which mark
the position of the groove in the rotor.
The screw for fastening the rotor must be tightened suffi-
ciently hand-tight so that the rotor does not come loose
during operation.
1
2
Fig. 21: Installing the rotor
1 Bottom side of rotor with groove
2 Rotor shaft with lateral driver
6.
When installing a rotor for the dual centrifugation,
note that it is imperative that the two rotor cavities be
positioned symmetrically (Fig. 22 ‘Rotor cavity sym-
metrically arranged’ on page 41). For this, it is prac-
tical to insert the adapter in the rotor cavities and to
align the rotor cavities with each other.
The rotor is installed in the opposite order.