en
Rev. 03 / 02.2021
AB3200
26 / 88
Commissioning
N > ROTOR MAX
PROG
T/°C
RCF
RPM
t/min
1
Fig. 12: Transport securing device
1 Transport securing device
3.
NOTICE
Before commissioning the device, remove the trans-
port securing device from the interior.
Remove the transport securing device.
4.
The device is now prepared for rotor installation.
If ‘Enter max cycles = <30000>’ is displayed after switching
on the device, the maximum permissible number of running
cycles (specified on the rotor) must first be entered before
the centrifugation run can be restarted.
If the speed in the activated program is higher than the
maximum rotor speed (Nmax), no centrifugation run can be