Safety information
Warning - Impermissible
ambient temperature
Warning - Position changed
due to vibration
Mechanical emergency
release
WARNING
Warning - Impermissible ambient temperature!
If the ambient temperature is outside of the permissible
ambient temperature range for the samples, these might be
compromised.
−
Observe the permissible maximum and minimum
ambient temperatures for setting up the device.
−
Do not place the device next to heat sources or devices
which radiate heat.
−
Check the samples after any device overheating to see if
they can still be used and take new samples, if neces-
sary.
−
Make sure the device is not exposed to direct sunlight.
−
Make sure the device is not exposed to frost.
−
Maintain the minimum clearance around the device.
−
Make sure that the fan is not blocked.
−
Never operate the device in a continuous run.
WARNING
Warning - Position changed due to vibration!
The device might fall down if the set-up surface vibrates or if
it is not flat or stable enough.
−
Only place the device on a stable, flat surface.
−
The set-up surface must reliably bear the weight of the
device.
WARNING
There is a risk of injury if the mechanical emergency release
of the device is actuated during the program run.
If the mechanical emergency release of the lid is actuated
during a program run, the program is aborted and the rotor
runs down unbraked.
−
Only actuate the emergency release when the rotor is
stopped.
−
Check whether the rotor has stopped before an emer-
gency release.