29 / 88
en
Rev. 03 / 02.2021
AB3200
Operation
T/°C
[Start-up and run-down parameters] key
■
Start-up levels, parameter
_/
Level 9 = shortest start-up time, ... level 1 = longest start-up time.
■
Start-up time, parameter
/
t. Adjustable in 1-second increments. The
adjustable time range depends on the set speed.
It is only possible to set start-up times when these are activated
Chapter 7.2.2 ‘Start-up and run-down parameters’ on page 48.
■
Braking levels, parameter
\_
1-9 = linear braking curve.
Level 9, = shortest run-down time, ... level 1 = long run-down time, level
0 = unbraked run-down.
■
Run-down time, parameter
\
t.
Adjustable in 1-second increments. The adjustable time range depends
on the set speed.
It is only possible to set run-down times when these are activated
Chapter 7.2.2 ‘Start-up and run-down parameters’ on page 48.
■
Braking switch-off speed; parameter N Brake.
Can be set from 50 RPM to the maximum rotor speed (Nmax), in incre-
ments of 10. After reaching this speed, the unbraked run-down occurs.
[T/°C] key
■
Can be set in degrees Celsius (°C) or in degrees Fahrenheit (°F). Setting
the temperature unit
Chapter 7.3.5.4 ‘Temp Unit’ on page 63.
■
Parameter
‘T/°C’ = degrees Celsius (°C).
Can be set from -20 °C to +40 °C, in increments of 1 °C.
■
Parameter
‘T/°F’ = degrees Fahrenheit (°F).
Can be set from -4 °F to +104 °F, in increments of 1 °F.
■
Parameter
‘T/°F’ = degrees Fahrenheit (°F).
Can be set from -4 °F to +194 °F, in increments of 1 °F.
■
The lowest reachable temperature depends on the rotor.
‘Rotors and accessories’ on page 86.
■
Scroll back in the menus.