Hameg HM 407A Manual Download Page 7

7

Subject to change without notice

Signal voltage U = 230Vrmsx2

2 = 651Vpp

(voltage > 160Vpp, with probe 10:1: U = 65.1Vpp),
desired display height H = min. 3.2div, max. 8div,
max. deflection coefficient D = 65.1:3.2 = 20.3V/div,
min. deflection coefficient D = 65.1:8 = 8.1V/div,
adjusted deflection coefficient D = 10V/div.

The previous examples are related to the CRT graticule reading.
The results can also be determined with the aid of the DV
cursor measurement (please note “controls and readout”).
The input voltage must not exceed 400V, independent from
the polarity.

If an AC voltage which is superimposed on a DC voltage is
applied, the maximum peak value of both voltages must not
or - 400V. So for AC voltages with a mean value of
zero volt the maximum peak to peak value is 800Vpp.

If attenuator probes with higher limits are used, the probes
limits are valid only if the oscilloscope is set to DC input
coupling.

If DC voltages are applied under AC input coupling conditions
the oscilloscope maximum input voltage value remains 400V.
The attenuator consists of a resistor in the probe and the 1M

input resistor of the oscilloscope, which are disabled by the AC
input coupling capacity when AC coupling is selected. This also
applies to DC voltages with superimposed AC voltages. It also
must be noted that due to the capacitive resistance of the AC
input coupling capacitor, the attenuation ratio depends on the
signal frequency. For sinewave signals with frequencies higher
than 40Hz this influence is negligible.

With the above listed exceptions

 HAMEG 

10:1 probes can be

used for DC measurements up to 600V or AC voltages (with
a mean value of zero volt) of 1200Vpp. The 100:1 probe

 HZ53

allows for 1200V DC or 2400Vpp for AC.

It should be noted that its AC peak value is derated at higher
frequencies. If a normal x10 probe is used to measure high
voltages there is the risk that the compensation trimmer
bridging the attenuator series resistor will break down causing
damage to the input of the oscilloscope. However, if for
example only the residual ripple of a high voltage is to be
displayed on the oscilloscope, a normal x10 probe is sufficient.
In this case, an appropriate high voltage capacitor (approx. 22-
68nF) must be connected in series with the input tip of the
probe.

With

 Y-POS. 

control (input coupling to

 GD

) it is possible to use

a horizontal graticule line as reference line for ground potential
before the measurement. It can lie below or above the horizon-
tal central line according to whether positive and/or negative
deviations from the ground potential are to be measured.

Total value of input voltage

The dotted line shows a voltage alternating at zero volt level.
If superimposed on a DC voltage, the addition of the positive

peak and the DC voltage results in the max. voltage (DC +
ACpeak).

Time Measurements

As a rule, most signals to be displayed are periodically repeating
processes, also called periods. The number of periods per
second is the repetition frequency. Depending on the time
base setting (TIME/DIV.-knob) indicated by the readout, one or
several signal periods or only a part of a period can be
displayed. The time coefficients are stated in ms/div, µs/div or
ns/div. The following examples are related to the CRT graticule
reading. The results can also be determined with the aid of the

T and 1/

T cursor measurement (please note “ controls and

readout”).

The duration of a signal period or a part of it is determined by
multiplying the relevant time (horizontal distance in div) by the
(calibrated) time coefficient displayed in the readout.
Uncalibrated, the time base speed can be reduced until a
maximum factor of 2.5 is reached. Therefore any intermediate
value is possible within the 1-2-5 sequence.

With the designations

L = displayed wave length in div of one period,
T = time in seconds for one period,
F = recurrence frequency in Hz of the signal,
Tc = time coefficient in ms, µs or ns/div and the relation
F = 1/T, the following equations can be stated:

However, these four values are not freely selectable. They
have to be within the following limits:

L

between 0.2 and 10div, if possible 4 to 10div,

T

between 10ns and 5s,

F

between 0.5Hz and 100MHz,

Tc

between 100ns/div and 500ms/div in 1-2-5 sequence
(with X-MAG. (x10) inactive), and

Tc

between 10ns/div and 50ms/div in 1-2-5 sequence
(with X-MAG. (x10) active).

Examples:
Displayed wavelength L = 7div,
set time coefficient Tc = 100ns/div,
required period T = 7x100x10

-9

 = 0.7µs

required rec. freq. F = 1:(0.7x10

-6

) = 1.428MHz.

Signal period T = 1s,
set time coefficient Tc = 0.2s/div,
required wavelength L = 1:0.2 = 5div.

Displayed ripple wavelength L = 1div,
set time coefficient Tc = 10ms/div,
required ripple freq. F = 1:(1x10x10

-3

) = 100Hz.

TV-line frequency F = 15625Hz,
set time coefficient Tc = 10µs/div,
required wavelength L = 1:(15 625x10

-5

) = 6.4div.

Sine wavelength L = min. 4div, max. 10div,
Frequency F = 1kHz,
max. time coefficient Tc = 1:(4x10

3

) = 0.25ms/div,

min. time coefficient Tc = 1:(10x10

3

) = 0.1ms/div,

set time coefficient Tc = 0.2ms/div,
required wavelength L = 1:(10

3

x0.2x10

-3

) = 5div.

Type of signal voltage

Summary of Contents for HM 407A

Page 1: ...in analog mode 29 Auto Set 31 Save Recall 31 Component Tester analog mode 31 General 31 Using the Component Tester 32 Test Procedure 32 Test Pattern Displays 32 Testing Resistors 32 Testing Capacitors...

Page 2: ...mity test made by HAMEG is based on the actual generic and product standards In cases where different limit values are applicable HAMEG applies the severer standard For emission the limits for residen...

Page 3: ...100MHz 0 5div Normal DC 100MHz LED for trigger indication Slope positive or negative Sources CH I or II line ext CH I alternate CH II 0 8div Coupling AC 10Hz 100MHz DC 0 100MHz HF 50kHz 100MHz LF 0 1...

Page 4: ...protective earth contact The protective action must not be negated by the use of an extension cord without a protective conductor Themains lineplugmustbeinsertedbeforeconnections are made to measuring...

Page 5: ...is stated for possible queries this helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain int...

Page 6: ...he readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general electrical engineering alternating voltage data normally refers to effective values rms root mean squ...

Page 7: ...ore the measurement It can lie below or above the horizon tal central line according to whether positive and or negative deviations from the ground potential are to be measured Total value of input vo...

Page 8: ...has little meaning For amplifiers with approximately constant group delay therefore good pulse transmission performance the following numerical relationship between rise time tr in ns and bandwidth B...

Page 9: ...tective conductor connections caused by external equipment connected to the mains line e g signal generators with interference protection capacitors Controls and readout The following description assu...

Page 10: ...a pre or post trigger value PT is not indicated by the readout Pressing and holding the STOR ON button switches over to the digital mode but without changing the channel operating mode CH I CH II DUAL...

Page 11: ...and back to PT0 The values refer to the X axis graticule of the screen display 10 1div The following description assumes that the X magnifier x10 is inactive and the signal display starts on the left...

Page 12: ...V In this case also several signal acquisition scans are required hence it is similar to Refresh operation The signal is averaged over the several acquisitions so that amplitude variations e g noise a...

Page 13: ...r pretrigger and post trigger settings the signal acquisition is not complete when the trigger occurs and will only be terminated later After termination the RES LED extinguishes but the signal displa...

Page 14: ...nob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to analog XY mode this control knob is inactive and the X POS knob must be used for a horizontal po...

Page 15: ...trigger point setting voltage The trigger unit starts the time base when the edge of a trigger signal crosses the trigger point In most Yt modes the trigger point is displayed in the readout by the s...

Page 16: ...nged All controls related to both channels are active if the inputs 31 and 35 are not set to GD 33 37 Whether alternated or chopped channel switching is present depends on the actual time base setting...

Page 17: ...ing in DUAL mode Under these conditions both TRIG CH I and CH II LEDs are lit As alternate triggering requires alternate channel operation alternate channel switching is set automatically A change of...

Page 18: ...of the trace can be set with the DEL POS control in the time base modes SEA SEARCH or DEL DELAY See SEA DEL ON OFF 29 STORAGE MODE ONLY The DEL POS control and the HO LED are disabled since in this m...

Page 19: ...eviously active settings Automatic normal triggering 15 trigger LEVEL 17 trigger slope 15 and trigger coupling 26 will be stored The instrument will be automatically switched to normal triggering NM a...

Page 20: ...as an earth ground symbol instead of the deflection coefficient and the AC or DC symbol The GD setting disables the input signal the AC DC 36 pushbutton and the VOLTS DIV 24 knob In automatic trigger...

Page 21: ...In this mode two signals are normally applied causing an X and a Y deflection The deflection coefficient selected for each channel may be different thus as in DUAL mode the V cursor measurement requi...

Page 22: ...ent the cursor moves fast 44 CAL Pushbutton and concentric socket A squarewave signal of 0 2Vpp 1 is available from the socket for probe adjustment purposes The signal frequency depends on the pushbut...

Page 23: ...eady for use If the AUTO SET function was not used and only a spot appears CAUTION CRT phosphor can be damaged reduce the intensity immediately and check that the XY mode is not selected XY not displa...

Page 24: ...displayed on the CRT screen should have the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator out...

Page 25: ...n the result The following must be noted here Becauseoftheperiodicnatureofthetrigonometricfunctions the calculation should be limited to angles 90 However here is the advantage of the method Due to ph...

Page 26: ...g signal AC voltage are displayable in Yt mode In this mode the signal voltage deflects the beam in vertical direction Y while the time base generator moves the beam from the left to the right of the...

Page 27: ...e previous slope setting for the undelayed time base trigger is stored and still active For further information please note Controls and readout The time base generator can be triggered by a rising or...

Page 28: ...lit 26 The trigger point symbol is inactive in line mains trigger mode as there is no direct amplitude relationship between the trigger voltage and the signal voltage A voltage originating from mains...

Page 29: ...sweep periods approx up to the ratio 10 1 Pulses or other signal waveforms appearing during this off period cannot trigger the time base Particularly with burst signals or aperiodic pulse trains of t...

Page 30: ...tion coefficient must be increased with TIME DIV knob A larger deflection coefficient than in the SEARCH mode cannot be set Example The SEARCH setting selected in figure 2 is 5ms cm The display in DEL...

Page 31: ...veral frequencies e g video signals are present AUTO SET sets the instrument automatically to the following operating conditions The input coupling is not affected while AC or DC is used but if GD was...

Page 32: ...e will tend towards the horizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can be approximately evaluated The determination of actua...

Page 33: ...n circuit tests are possible in many cases However they are not well defined This is caused by a shunt connection of real or complex impedances especially if they are of relatively low impedance at 50...

Page 34: ...rizontal center graticule line by using the Y POS control Then the trigger point symbol may be set 2 division above the 0 Volt position using the LEVEL control If the deflection coefficient is set to...

Page 35: ...commended to switch on the instrument for about 20 minutes prior to the commencement of any check Cathode Ray Tube Brightness and Focus Linearity Raster Distortion Normally the CRT of the instrument h...

Page 36: ...regular intervals the deflection coefficients on all positions of the input attenuators and readjust them as necessary A compensated 2 1 series attenuator is also necessary and this must be matched t...

Page 37: ...For precise determination set the peak of the first marker or cycle peak exactly behind the first vertical graticule line using the X POS control Deviation tendencies can be noted after some of the m...

Page 38: ...ariable resistor 100k located on the CRT PCB is used for this adjustment procedure It may only be touched by a properlyinsulatingscrewdriver Caution Highvoltage Correct adjustment is achieved when the...

Page 39: ...a high level reference potential is not permitted and endangers operator oscilloscope interface and peripheral devices In case of disregard of the safety warnings contained in this manual HAMEG refuse...

Page 40: ...40 Subject to change without notice Front Panel HM407...

Reviews: