Hameg HM 407A Manual Download Page 5

5

Subject to change without notice

explosive, corrosive, dusty, or moist environments. The
oscilloscope can be operated in any position, but the convection
cooling must not be impaired. The ventilation holes may not be
covered. For continuous operation the instrument should be
used in the horizontal position, preferably tilted upwards,
resting on the tilt handle.

The specifications stating tolerances are only valid if
the instrument has warmed up for 30minutes at an
ambient temperature b15°C (+59°F) and +30°C
(+86°F). Values without tolerances are typical for an
average instrument.

EMC

This instrument conforms to the European standards regarding
the electromagnetic compatibility. The applied standards are:
Generic immunity standard EN50082-2:1995 (for industrial
environment) Generic emission standard EN50081-1:1992
(for residential, commercial and light industry environment).

This means that the instrument has been tested to the highest
standards.

Please note that under the influence of strong electromagnetic
fields, such signals may be superimposed on the measured
signals.

Under certain conditions this is  unavoidable due to the
instrument’s high input sensitivity, high input impedance and
bandwidth. Shielded measuring cables, shielding and earthing
of the device under test may reduce or eliminate those effects.

Warranty

HAMEG warrants to its Customers that the products it
manufactures and sells will be free from defects in materials
and workmanship for a period of 2 years. This warranty shall
not apply to any defect, failure or damage caused by improper
use or inadequate maintenance and care. HAMEG shall not be
obliged to provide service under this warranty to repair damage
resulting from attempts by personnel other than HAMEG
representatives to install, repair, service or modify these
products.

In order to obtain service under this warranty, Customers
must contact and notify the distributor who has sold the
product. Each instrument is subjected to a quality test with 10
hour burn-in before leaving the production. Practically all early
failures are detected by this method. In the case of shipments
by post, rail or carrier it is recommended that the original
packing is carefully preserved. Transport damages and damage
due to gross negligence are not covered by the guarantee.

In the case of a complaint, a label should be attached to the
housing of the instrument which describes briefly the faults
observed. If at the same time the name and telephone number
(dialing code and telephone or direct number or department
designation) is stated for possible queries, this helps towards
speeding up the processing of guarantee claims.

Maintenance

Various important properties of the oscilloscope should be
carefully checked at certain intervals. Only in this way is it
largely certain that all signals are displayed with the accuracy
on which the technical data are based. The test methods
described in the test plan of this manual can be performed
without great expenditure on measuring instruments. However,
purchase of the HAMEG scope tester HZ 60, which despite its
low price is highly suitable for tasks of this type, is very much

recommended. The exterior of the oscilloscope should be
cleaned regularly with a dusting brush. Dirt which is difficult to
remove on the casing and handle, the plastic and aluminium
parts, can be removed with a moistened cloth (99% water
+1% mild detergent). Spirit or washing benzine (petroleum
ether) can be used to remove greasy dirt. The screen may be
cleaned with water or washing benzine (but not with spirit
(alcohol) or solvents), it must then be wiped with a dry clean
lint-free cloth. Under no circumstances may the cleaning fluid
get into the instrument. The use of other cleaning agents can
attack the plastic and paint surfaces.

Protective Switch-Off

This instrument is equipped with a switch mode power supply.
It has both overvoltage and overload protection, which will
cause the switch mode supply to limit power consumption to
a minimum. In this case a ticking noise may be heard.

Power supply

The oscilloscope operates on mains/line voltages between
100VAC and 240VAC. No means of switching to different input
voltages has therefore been provided.

The power input fuses are externally accessible. The fuseholder
is located above the 3-pole power connector. The power input
fuses are externally accessible, if the rubber connector is
removed. The fuseholder can be released by pressing its
plastic retainers with the aid of a small screwdriver. The
retainers are located on the right and left side of the holder and
must be pressed towards the center. The fuse(s) can then be
replaced and pressed in until locked on both sides.

Use of patched fuses or short-circuiting of the fuseholder is
not permissible; HAMEG assumes no liability whatsoever for
any damage caused as a result, and all warranty claims
become null and void.

Fuse type:
Size 5x20mm; 0.8A, 250V AC fuse;
must meet IEC specification 127,
Sheet III (or DIN 41 662
or DIN 41 571, sheet 3).
Time characteristic: time-lag (T).

Attention!
There is a fuse located inside the instrument within the
switch mode power supply:

Size 5x20mm; 0.8A, 250V AC fuse;
must meet IEC specification 127,
Sheet III (or DIN 41 662
or DIN 41 571, sheet 3).
Time characteristic: fast (F).

This fuse must not be replaced by the operator!

General Information

Summary of Contents for HM 407A

Page 1: ...in analog mode 29 Auto Set 31 Save Recall 31 Component Tester analog mode 31 General 31 Using the Component Tester 32 Test Procedure 32 Test Pattern Displays 32 Testing Resistors 32 Testing Capacitors...

Page 2: ...mity test made by HAMEG is based on the actual generic and product standards In cases where different limit values are applicable HAMEG applies the severer standard For emission the limits for residen...

Page 3: ...100MHz 0 5div Normal DC 100MHz LED for trigger indication Slope positive or negative Sources CH I or II line ext CH I alternate CH II 0 8div Coupling AC 10Hz 100MHz DC 0 100MHz HF 50kHz 100MHz LF 0 1...

Page 4: ...protective earth contact The protective action must not be negated by the use of an extension cord without a protective conductor Themains lineplugmustbeinsertedbeforeconnections are made to measuring...

Page 5: ...is stated for possible queries this helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain int...

Page 6: ...he readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general electrical engineering alternating voltage data normally refers to effective values rms root mean squ...

Page 7: ...ore the measurement It can lie below or above the horizon tal central line according to whether positive and or negative deviations from the ground potential are to be measured Total value of input vo...

Page 8: ...has little meaning For amplifiers with approximately constant group delay therefore good pulse transmission performance the following numerical relationship between rise time tr in ns and bandwidth B...

Page 9: ...tective conductor connections caused by external equipment connected to the mains line e g signal generators with interference protection capacitors Controls and readout The following description assu...

Page 10: ...a pre or post trigger value PT is not indicated by the readout Pressing and holding the STOR ON button switches over to the digital mode but without changing the channel operating mode CH I CH II DUAL...

Page 11: ...and back to PT0 The values refer to the X axis graticule of the screen display 10 1div The following description assumes that the X magnifier x10 is inactive and the signal display starts on the left...

Page 12: ...V In this case also several signal acquisition scans are required hence it is similar to Refresh operation The signal is averaged over the several acquisitions so that amplitude variations e g noise a...

Page 13: ...r pretrigger and post trigger settings the signal acquisition is not complete when the trigger occurs and will only be terminated later After termination the RES LED extinguishes but the signal displa...

Page 14: ...nob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to analog XY mode this control knob is inactive and the X POS knob must be used for a horizontal po...

Page 15: ...trigger point setting voltage The trigger unit starts the time base when the edge of a trigger signal crosses the trigger point In most Yt modes the trigger point is displayed in the readout by the s...

Page 16: ...nged All controls related to both channels are active if the inputs 31 and 35 are not set to GD 33 37 Whether alternated or chopped channel switching is present depends on the actual time base setting...

Page 17: ...ing in DUAL mode Under these conditions both TRIG CH I and CH II LEDs are lit As alternate triggering requires alternate channel operation alternate channel switching is set automatically A change of...

Page 18: ...of the trace can be set with the DEL POS control in the time base modes SEA SEARCH or DEL DELAY See SEA DEL ON OFF 29 STORAGE MODE ONLY The DEL POS control and the HO LED are disabled since in this m...

Page 19: ...eviously active settings Automatic normal triggering 15 trigger LEVEL 17 trigger slope 15 and trigger coupling 26 will be stored The instrument will be automatically switched to normal triggering NM a...

Page 20: ...as an earth ground symbol instead of the deflection coefficient and the AC or DC symbol The GD setting disables the input signal the AC DC 36 pushbutton and the VOLTS DIV 24 knob In automatic trigger...

Page 21: ...In this mode two signals are normally applied causing an X and a Y deflection The deflection coefficient selected for each channel may be different thus as in DUAL mode the V cursor measurement requi...

Page 22: ...ent the cursor moves fast 44 CAL Pushbutton and concentric socket A squarewave signal of 0 2Vpp 1 is available from the socket for probe adjustment purposes The signal frequency depends on the pushbut...

Page 23: ...eady for use If the AUTO SET function was not used and only a spot appears CAUTION CRT phosphor can be damaged reduce the intensity immediately and check that the XY mode is not selected XY not displa...

Page 24: ...displayed on the CRT screen should have the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator out...

Page 25: ...n the result The following must be noted here Becauseoftheperiodicnatureofthetrigonometricfunctions the calculation should be limited to angles 90 However here is the advantage of the method Due to ph...

Page 26: ...g signal AC voltage are displayable in Yt mode In this mode the signal voltage deflects the beam in vertical direction Y while the time base generator moves the beam from the left to the right of the...

Page 27: ...e previous slope setting for the undelayed time base trigger is stored and still active For further information please note Controls and readout The time base generator can be triggered by a rising or...

Page 28: ...lit 26 The trigger point symbol is inactive in line mains trigger mode as there is no direct amplitude relationship between the trigger voltage and the signal voltage A voltage originating from mains...

Page 29: ...sweep periods approx up to the ratio 10 1 Pulses or other signal waveforms appearing during this off period cannot trigger the time base Particularly with burst signals or aperiodic pulse trains of t...

Page 30: ...tion coefficient must be increased with TIME DIV knob A larger deflection coefficient than in the SEARCH mode cannot be set Example The SEARCH setting selected in figure 2 is 5ms cm The display in DEL...

Page 31: ...veral frequencies e g video signals are present AUTO SET sets the instrument automatically to the following operating conditions The input coupling is not affected while AC or DC is used but if GD was...

Page 32: ...e will tend towards the horizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can be approximately evaluated The determination of actua...

Page 33: ...n circuit tests are possible in many cases However they are not well defined This is caused by a shunt connection of real or complex impedances especially if they are of relatively low impedance at 50...

Page 34: ...rizontal center graticule line by using the Y POS control Then the trigger point symbol may be set 2 division above the 0 Volt position using the LEVEL control If the deflection coefficient is set to...

Page 35: ...commended to switch on the instrument for about 20 minutes prior to the commencement of any check Cathode Ray Tube Brightness and Focus Linearity Raster Distortion Normally the CRT of the instrument h...

Page 36: ...regular intervals the deflection coefficients on all positions of the input attenuators and readjust them as necessary A compensated 2 1 series attenuator is also necessary and this must be matched t...

Page 37: ...For precise determination set the peak of the first marker or cycle peak exactly behind the first vertical graticule line using the X POS control Deviation tendencies can be noted after some of the m...

Page 38: ...ariable resistor 100k located on the CRT PCB is used for this adjustment procedure It may only be touched by a properlyinsulatingscrewdriver Caution Highvoltage Correct adjustment is achieved when the...

Page 39: ...a high level reference potential is not permitted and endangers operator oscilloscope interface and peripheral devices In case of disregard of the safety warnings contained in this manual HAMEG refuse...

Page 40: ...40 Subject to change without notice Front Panel HM407...

Reviews: