Hameg HM 407A Manual Download Page 34

34

Subject to change without notice

enables the user to select the most suitable mode for the
specific measurement. The

 HM407 

contains two 8 bit flash A/

D converters with a maximum sampling rate of 50MS/s each.
Except in the combination of

 DUAL 

mode and

 SINGLE 

event

recording with a maximum of 40MS/s (time coefficient 5µs/
div) for each channel, in all other modes 100MS/s is available
if the lowest time coefficient is chosen. Apart from the factors
above, there is no principle difference between capturing
repetitive signals and one-time events. The signal is always
displayed with a linear connection between the dots (Dot Join
function).

Signals captured and stored in storage mode can be called via
the built-in RS232 interface for documentation purposes. For
further information please note section “RS232 Interface -
Remote Control”.

Signal recording modes

Signals can be recorded and displayed in six different modes:

REFRESH 

mode (RFR LED lit, readout indicates RFR)

ENVELOPE 

mode (ENV LED lit, readout indicates ENV)

AVERAGE 

mode (AVM LED lit, readout indicates AV...)

SINGLE 

mode (SGL LED lit, readout indicates SGL)

ROLL 

mode (ROL LED lit, readout indicates ROL)

XY 

mode (RFR LED lit, readout indicates XY and the sampling rate)

Except

 ROLL 

and

 XY 

mode, a signal recording in all other

modes requires a trigger signal.

In

 REFRESH, ENVELOPE 

and

 AVERAGE 

modes the

instruments behaves like an analog oscilloscope. The trigger
circuit starts a recording, overwriting the previous recording
from the left to the right side of the screen. After the recording
has been finished, the next trigger event starts the same
procedure. This can also be caused in automatic trigger mode
without an applied signal by the automatic circuitry. Then only
the trace (

Y-POS. 

setting)  is recorded.

In contrast to automatic trigger mode, in normal trigger mode
the automatic system is switched off and consequently only
a trigger signal can start a recording. Unlike analog mode
where the screen is dark until a trigger signal starts the time
base, in store mode the last recorded signal remains visible as
long as no new recording is triggered by an input signal.

AVERAGE 

and

 ENVELOPE 

are

 REFRESH 

sub-modes and

described in section “Controls and readout” under item

 (9)

.

SINGLE 

mode

 (SGL) 

enables the capture of one-time events,

started by a suitable trigger signal. It is recommended, to
select input ground

 (GD) 

condition and set the trace on a

graticule line which is then used as the 0 Volt position
(reference) line. In

 SINGLE 

mode the instrument is

automatically set to normal trigger mode. After

 SINGLE (SGL)

mode has been selected, the trigger point symbol should be
set above or below the 0 Volt position line, according to the
expected voltage of the event to be captured. Whether the
slope selection is set for a rising or falling slope depends on the
measurement task. After this procedure AC or DC input
coupling must be selected and the signal capture is started
after activating the RESET function. The following example is
for additional explanation.

If a voltage drop on a +5 Volt line is expected, the 0 Volt line
may be set to the horizontal center graticule line by using the

Y-POS. 

control. Then the trigger point symbol may be set 2

division above the 0 Volt position using the

 LEVEL 

control. If

the deflection coefficient is set to 200mV and

 DC 

input

coupling selected, using a x10 probe, the trigger point is at +4
Volts. A voltage drop from +5 Volts below +4 Volts then is used

for triggering, if negative slope setting is selected. If the slope
selection is positive the end of the event (voltage drop ended)
is used for triggering.

For explanations regarding ROLL mode, please note this item

(9) 

in section “Readout and Controls”.

Vertical resolution

The dot density in each operation mode is 8 bits = 2

8

 = 256 dots

displayed over a height of roughly 10 divisions. The instrument
is adjusted for 25 dots per division.
This eases processing and cursor measurement. Insignificant
differences between the (analog) screen display and the
(digital) data are unavoidable.

This concerns signal height as well as the position. The trace
position is defined in respect to the following horizontal
graticule lines:

Center line

= 10000000 (binary) = 80 (hex)

= 128 (dec).

Top line

= 11100100 (binary) = E4 (hex)

= 228 (dec).

Bottom line = 00011100 (binary) = 1C (hex) = 28 (dec).

In contrast to analog mode with its theoretically unlimited
resolution, the vertical resolution has 25 possible trace positions
per division.

If the signal is superimposed by noise or a critical Y-POS.
setting is used, the least significant bit (LSB) may change
continuously. This additionally reduces the vertical resolution
in storage mode, but is unavoidable. In contrast to the expen-
sive flash A/D converters used in this instrument, other
converters such as CCD cause more noise.

Horizontal resolution

The maximum number of signals to be displayed simultaneously
is three. Each signal consists of 2048 (2

11

) byte (samples).

Referred to the horizontal raster, the resolution is 200 samples
per division.

Pure (only) digital oscilloscope with VGA monitor type CRTs
offer only 50 samples per division. If LCD displays are used the
current resolution is 25 samples per division. For a given time
base setting the

 HAMEG 

instrument samples at a 4 (compared

to VGA) or 8 ( referred to LCD) times higher sampling rate. The
higher number of samples/div results in a shorter sampling
interval. For the following example it must be kept in mind, that
the time base setting is related to the signal period duration and
consequently should enable the display of one complete signal
period. If e.g. a 50Hz signal has to be displayed the time base
should be set to 2ms/div. The maximum signal frequency of a
superimposed sinewave signal, which must be sampled with at
least 10 samples per period, depends on the horizontal resolution:

samples/div

sampling interval

sampling rate max frequency

200

2ms : 200 = 10µs

100kS/s

10kHz

50

2ms : 50 = 40µs

25kS/s

2.5kHz

25

2ms : 25 = 80µs

2.5kS/s

1.25kHz

During X magnifier mode the record length is still 2048 byte.
The CRT displays a tenth of the recording (20 byte/division)
plus 180 byte/div calculated in linear interpolation operation by
a RISC processor. The smallest available time coefficient then
is 50ns/div instead of 500ns/div without the X magnifier.

Maximum signal frequency in storage mode

The highest capturable signal frequency cannot be exactly
defined, since it depends to a large extent on the waveform.

Storage Mode

Summary of Contents for HM 407A

Page 1: ...in analog mode 29 Auto Set 31 Save Recall 31 Component Tester analog mode 31 General 31 Using the Component Tester 32 Test Procedure 32 Test Pattern Displays 32 Testing Resistors 32 Testing Capacitors...

Page 2: ...mity test made by HAMEG is based on the actual generic and product standards In cases where different limit values are applicable HAMEG applies the severer standard For emission the limits for residen...

Page 3: ...100MHz 0 5div Normal DC 100MHz LED for trigger indication Slope positive or negative Sources CH I or II line ext CH I alternate CH II 0 8div Coupling AC 10Hz 100MHz DC 0 100MHz HF 50kHz 100MHz LF 0 1...

Page 4: ...protective earth contact The protective action must not be negated by the use of an extension cord without a protective conductor Themains lineplugmustbeinsertedbeforeconnections are made to measuring...

Page 5: ...is stated for possible queries this helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain int...

Page 6: ...he readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general electrical engineering alternating voltage data normally refers to effective values rms root mean squ...

Page 7: ...ore the measurement It can lie below or above the horizon tal central line according to whether positive and or negative deviations from the ground potential are to be measured Total value of input vo...

Page 8: ...has little meaning For amplifiers with approximately constant group delay therefore good pulse transmission performance the following numerical relationship between rise time tr in ns and bandwidth B...

Page 9: ...tective conductor connections caused by external equipment connected to the mains line e g signal generators with interference protection capacitors Controls and readout The following description assu...

Page 10: ...a pre or post trigger value PT is not indicated by the readout Pressing and holding the STOR ON button switches over to the digital mode but without changing the channel operating mode CH I CH II DUAL...

Page 11: ...and back to PT0 The values refer to the X axis graticule of the screen display 10 1div The following description assumes that the X magnifier x10 is inactive and the signal display starts on the left...

Page 12: ...V In this case also several signal acquisition scans are required hence it is similar to Refresh operation The signal is averaged over the several acquisitions so that amplitude variations e g noise a...

Page 13: ...r pretrigger and post trigger settings the signal acquisition is not complete when the trigger occurs and will only be terminated later After termination the RES LED extinguishes but the signal displa...

Page 14: ...nob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to analog XY mode this control knob is inactive and the X POS knob must be used for a horizontal po...

Page 15: ...trigger point setting voltage The trigger unit starts the time base when the edge of a trigger signal crosses the trigger point In most Yt modes the trigger point is displayed in the readout by the s...

Page 16: ...nged All controls related to both channels are active if the inputs 31 and 35 are not set to GD 33 37 Whether alternated or chopped channel switching is present depends on the actual time base setting...

Page 17: ...ing in DUAL mode Under these conditions both TRIG CH I and CH II LEDs are lit As alternate triggering requires alternate channel operation alternate channel switching is set automatically A change of...

Page 18: ...of the trace can be set with the DEL POS control in the time base modes SEA SEARCH or DEL DELAY See SEA DEL ON OFF 29 STORAGE MODE ONLY The DEL POS control and the HO LED are disabled since in this m...

Page 19: ...eviously active settings Automatic normal triggering 15 trigger LEVEL 17 trigger slope 15 and trigger coupling 26 will be stored The instrument will be automatically switched to normal triggering NM a...

Page 20: ...as an earth ground symbol instead of the deflection coefficient and the AC or DC symbol The GD setting disables the input signal the AC DC 36 pushbutton and the VOLTS DIV 24 knob In automatic trigger...

Page 21: ...In this mode two signals are normally applied causing an X and a Y deflection The deflection coefficient selected for each channel may be different thus as in DUAL mode the V cursor measurement requi...

Page 22: ...ent the cursor moves fast 44 CAL Pushbutton and concentric socket A squarewave signal of 0 2Vpp 1 is available from the socket for probe adjustment purposes The signal frequency depends on the pushbut...

Page 23: ...eady for use If the AUTO SET function was not used and only a spot appears CAUTION CRT phosphor can be damaged reduce the intensity immediately and check that the XY mode is not selected XY not displa...

Page 24: ...displayed on the CRT screen should have the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator out...

Page 25: ...n the result The following must be noted here Becauseoftheperiodicnatureofthetrigonometricfunctions the calculation should be limited to angles 90 However here is the advantage of the method Due to ph...

Page 26: ...g signal AC voltage are displayable in Yt mode In this mode the signal voltage deflects the beam in vertical direction Y while the time base generator moves the beam from the left to the right of the...

Page 27: ...e previous slope setting for the undelayed time base trigger is stored and still active For further information please note Controls and readout The time base generator can be triggered by a rising or...

Page 28: ...lit 26 The trigger point symbol is inactive in line mains trigger mode as there is no direct amplitude relationship between the trigger voltage and the signal voltage A voltage originating from mains...

Page 29: ...sweep periods approx up to the ratio 10 1 Pulses or other signal waveforms appearing during this off period cannot trigger the time base Particularly with burst signals or aperiodic pulse trains of t...

Page 30: ...tion coefficient must be increased with TIME DIV knob A larger deflection coefficient than in the SEARCH mode cannot be set Example The SEARCH setting selected in figure 2 is 5ms cm The display in DEL...

Page 31: ...veral frequencies e g video signals are present AUTO SET sets the instrument automatically to the following operating conditions The input coupling is not affected while AC or DC is used but if GD was...

Page 32: ...e will tend towards the horizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can be approximately evaluated The determination of actua...

Page 33: ...n circuit tests are possible in many cases However they are not well defined This is caused by a shunt connection of real or complex impedances especially if they are of relatively low impedance at 50...

Page 34: ...rizontal center graticule line by using the Y POS control Then the trigger point symbol may be set 2 division above the 0 Volt position using the LEVEL control If the deflection coefficient is set to...

Page 35: ...commended to switch on the instrument for about 20 minutes prior to the commencement of any check Cathode Ray Tube Brightness and Focus Linearity Raster Distortion Normally the CRT of the instrument h...

Page 36: ...regular intervals the deflection coefficients on all positions of the input attenuators and readjust them as necessary A compensated 2 1 series attenuator is also necessary and this must be matched t...

Page 37: ...For precise determination set the peak of the first marker or cycle peak exactly behind the first vertical graticule line using the X POS control Deviation tendencies can be noted after some of the m...

Page 38: ...ariable resistor 100k located on the CRT PCB is used for this adjustment procedure It may only be touched by a properlyinsulatingscrewdriver Caution Highvoltage Correct adjustment is achieved when the...

Page 39: ...a high level reference potential is not permitted and endangers operator oscilloscope interface and peripheral devices In case of disregard of the safety warnings contained in this manual HAMEG refuse...

Page 40: ...40 Subject to change without notice Front Panel HM407...

Reviews: