vernier (variable). The current setting is displayed by the
VAR-LED located above the knob. After switching the

VAR-LED (24) 

on, the deflection coefficient is still

calibrated. Turning the

 VOLTS/DIV. (24)

 control knob

counter clockwise reduces the signal height and the
deflection coefficient becomes uncalibrated.

The readout then displays i.e.

 “Y2>...” 

indicating the

uncalibrated condition instead of

 “Y2:...”

. Pressing and

holding the CH II pushbutton again switches the LED off,
sets the deflection coefficient into calibrated condition
and activates the attenuator function. The previous vernier
setting will not be stored.

The

 CH II 

pushbutton can also be pressed simultaneously

with the

 DUAL(22) 

button.

 Please note item (22).

(26) TRIG. MODE

Pushbuttons and associated indicator LEDs.

Pressing the upper or lower button selects the trigger
coupling. The actual setting is indicated by 

a TRIG.

MODE-LED (26)

.

Each time the lower

 TRIG. MODE 

pushbutton is pressed

the trigger coupling changes in the sequence:

AC

DC content suppressed,

DC

DC content effective (peak value detection inactive),

HF

high-pass filter cuts off frequencies below
approx. 50kHz (trigger point symbol switched off),

LF

low-pass filter cuts off frequencies above
approx. 1.5kHz, in combination with automatic
triggering AC or normal triggering DC coupled,

TVL

TV signal, line pulse triggering,
trigger point symbol switched off,

TVF

TV signal, frame pulse triggering,
trigger point symbol switched off.

~

line/mains triggering, trigger point symbol
and TRIG. LED (23) are switched off.

In some trigger modes such as alternate triggering, some
trigger coupling modes are automatically disabled and
can not be selected.

(27) DEL.POS.

Rotary knob with two functions and related  HO-LED.

The

 DEL.POS. 

knob functions as a hold off time control,

when the time base is not working in the

 SEA. (SEARCH)

or in

 DEL. (DELAY) 

mode. The HO-LED is not lit when the

hold off time is set to minimum. The HO-LED lights up and
the hold off time increases as the knob is rotated clockwise.
A signal sounds on reaching the maximum hold off time.
Similarly in the opposite direction until minimum hold off
time is reached (HO-LED extinguishes). The hold off time

is automatically set to minimum when the time base is
changed. (For the application of hold off time setting see
the paragraph with the same heading).

The start of the trace can be set with the

 DEL.POS.

control in the time base modes

 SEA. (SEARCH) 

or

 DEL.

(DELAY)

. See  SEA. / DEL. - ON / OFF (29).

STORAGE MODE ONLY

The

 DEL.POS. 

control and the HO-LED are disabled,

since in this mode the hold off  time is set to minimum.
The hold off time last used in the analog mode is not
saved. Consequently, the hold off time is set to minimum
when it is switched back to analog mode.

The

 DEL.POS. 

control cannot be used for delay setting

since neither

 SEA. (SEARCH) 

nor

 DEL. (DELAY) 

time

base operation is available.

(28) TIME/DIV. 

- Control knob.

The time base is set with this knob in the TIME/DIV. field,
and the setting is displayed at the top left in the readout
(e.g. ”T:10µs”). This knob acts as the time base step
switch when the VAR-LED above it is not lit. Then, the
time deflection coefficient can be set in a 1-2-5 sequence
and the time base is calibrated. Rotating anticlockwise
increases the deflection coefficient and  rotating clockwise
decreases the deflection. The control acts as a vernier
(fine adjustment) when the VAR-LED is lit.

The following description refers to the function as a time
base switch.

Attention
The different deflection coefficient ranges between the
analog and digital time bases require special conside-
ration when switching between analog and digital
modes. These are described in section  (7).

ANALOG MODE ONLY

Time deflection coefficients between  500ms/div. and
50ns/div. in a 1-2-5 sequence  can be selected without X-
MAG. x10 magnification. Time delay between 120ms and
200ns can be selected in the

 ”SEA” (SEARCH) 

mode.

Time deflection coefficient range in the

  ”DEL” (DELAY)

mode extends from 20ms/div. to 50ns/div.

STORAGE MODE ONLY

The time base can be set to deflection coefficients
between 100s/div. and 1µs/div in  storage mode, if X-
MAG x10 is inactive.

(29) SEA./DEL. - ON/OFF  

pushbutton

The pushbutton is operative only in analog mode.

This button is used to switch between delayed and undelayed
time base. The delayed time base operation enables a
magnified display in X-direction which is otherwise only
possible with a second time base. Pressing and holding the
button switches from normal time base mode to

 SEA.

(SEARCH)

, when currently neither

  ”SEA” (SEARCH)  

nor

”DEL” (DELAY) 

operation or

 ”DTR” 

(triggered DELAY) is

effective. Afterwards, it can be switched between

 SEA. 

and

DEL. 

by briefly pressing the button.

These operating modes are indicated in the readout to the
right of the trigger slope indication thus:
In case of  SEARCH , ”SEA” will be displayed;
In untriggered DELAY mode, ”DEL” (DEL.) and
in  triggered  DELAY mode, ”DTR” (DEL.TRIG.).

Controls and readout

Summary of Contents for HM 407A

Page 1: ...in analog mode 29 Auto Set 31 Save Recall 31 Component Tester analog mode 31 General 31 Using the Component Tester 32 Test Procedure 32 Test Pattern Displays 32 Testing Resistors 32 Testing Capacitors...

Page 2: ...mity test made by HAMEG is based on the actual generic and product standards In cases where different limit values are applicable HAMEG applies the severer standard For emission the limits for residen...

Page 3: ...100MHz 0 5div Normal DC 100MHz LED for trigger indication Slope positive or negative Sources CH I or II line ext CH I alternate CH II 0 8div Coupling AC 10Hz 100MHz DC 0 100MHz HF 50kHz 100MHz LF 0 1...

Page 4: ...protective earth contact The protective action must not be negated by the use of an extension cord without a protective conductor Themains lineplugmustbeinsertedbeforeconnections are made to measuring...

Page 5: ...is stated for possible queries this helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain int...

Page 6: ...he readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general electrical engineering alternating voltage data normally refers to effective values rms root mean squ...

Page 7: ...ore the measurement It can lie below or above the horizon tal central line according to whether positive and or negative deviations from the ground potential are to be measured Total value of input vo...

Page 8: ...has little meaning For amplifiers with approximately constant group delay therefore good pulse transmission performance the following numerical relationship between rise time tr in ns and bandwidth B...

Page 9: ...tective conductor connections caused by external equipment connected to the mains line e g signal generators with interference protection capacitors Controls and readout The following description assu...

Page 10: ...a pre or post trigger value PT is not indicated by the readout Pressing and holding the STOR ON button switches over to the digital mode but without changing the channel operating mode CH I CH II DUAL...

Page 11: ...and back to PT0 The values refer to the X axis graticule of the screen display 10 1div The following description assumes that the X magnifier x10 is inactive and the signal display starts on the left...

Page 12: ...V In this case also several signal acquisition scans are required hence it is similar to Refresh operation The signal is averaged over the several acquisitions so that amplitude variations e g noise a...

Page 13: ...r pretrigger and post trigger settings the signal acquisition is not complete when the trigger occurs and will only be terminated later After termination the RES LED extinguishes but the signal displa...

Page 14: ...nob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to analog XY mode this control knob is inactive and the X POS knob must be used for a horizontal po...

Page 15: ...trigger point setting voltage The trigger unit starts the time base when the edge of a trigger signal crosses the trigger point In most Yt modes the trigger point is displayed in the readout by the s...

Page 16: ...nged All controls related to both channels are active if the inputs 31 and 35 are not set to GD 33 37 Whether alternated or chopped channel switching is present depends on the actual time base setting...

Page 17: ...ing in DUAL mode Under these conditions both TRIG CH I and CH II LEDs are lit As alternate triggering requires alternate channel operation alternate channel switching is set automatically A change of...

Page 18: ...of the trace can be set with the DEL POS control in the time base modes SEA SEARCH or DEL DELAY See SEA DEL ON OFF 29 STORAGE MODE ONLY The DEL POS control and the HO LED are disabled since in this m...

Page 19: ...eviously active settings Automatic normal triggering 15 trigger LEVEL 17 trigger slope 15 and trigger coupling 26 will be stored The instrument will be automatically switched to normal triggering NM a...

Page 20: ...as an earth ground symbol instead of the deflection coefficient and the AC or DC symbol The GD setting disables the input signal the AC DC 36 pushbutton and the VOLTS DIV 24 knob In automatic trigger...

Page 21: ...In this mode two signals are normally applied causing an X and a Y deflection The deflection coefficient selected for each channel may be different thus as in DUAL mode the V cursor measurement requi...

Page 22: ...ent the cursor moves fast 44 CAL Pushbutton and concentric socket A squarewave signal of 0 2Vpp 1 is available from the socket for probe adjustment purposes The signal frequency depends on the pushbut...

Page 23: ...eady for use If the AUTO SET function was not used and only a spot appears CAUTION CRT phosphor can be damaged reduce the intensity immediately and check that the XY mode is not selected XY not displa...

Page 24: ...displayed on the CRT screen should have the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator out...

Page 25: ...n the result The following must be noted here Becauseoftheperiodicnatureofthetrigonometricfunctions the calculation should be limited to angles 90 However here is the advantage of the method Due to ph...

Page 26: ...g signal AC voltage are displayable in Yt mode In this mode the signal voltage deflects the beam in vertical direction Y while the time base generator moves the beam from the left to the right of the...

Page 27: ...e previous slope setting for the undelayed time base trigger is stored and still active For further information please note Controls and readout The time base generator can be triggered by a rising or...

Page 28: ...lit 26 The trigger point symbol is inactive in line mains trigger mode as there is no direct amplitude relationship between the trigger voltage and the signal voltage A voltage originating from mains...

Page 29: ...sweep periods approx up to the ratio 10 1 Pulses or other signal waveforms appearing during this off period cannot trigger the time base Particularly with burst signals or aperiodic pulse trains of t...

Page 30: ...tion coefficient must be increased with TIME DIV knob A larger deflection coefficient than in the SEARCH mode cannot be set Example The SEARCH setting selected in figure 2 is 5ms cm The display in DEL...

Page 31: ...veral frequencies e g video signals are present AUTO SET sets the instrument automatically to the following operating conditions The input coupling is not affected while AC or DC is used but if GD was...

Page 32: ...e will tend towards the horizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can be approximately evaluated The determination of actua...

Page 33: ...n circuit tests are possible in many cases However they are not well defined This is caused by a shunt connection of real or complex impedances especially if they are of relatively low impedance at 50...

Page 34: ...rizontal center graticule line by using the Y POS control Then the trigger point symbol may be set 2 division above the 0 Volt position using the LEVEL control If the deflection coefficient is set to...

Page 35: ...commended to switch on the instrument for about 20 minutes prior to the commencement of any check Cathode Ray Tube Brightness and Focus Linearity Raster Distortion Normally the CRT of the instrument h...

Page 36: ...regular intervals the deflection coefficients on all positions of the input attenuators and readjust them as necessary A compensated 2 1 series attenuator is also necessary and this must be matched t...

Page 37: ...For precise determination set the peak of the first marker or cycle peak exactly behind the first vertical graticule line using the X POS control Deviation tendencies can be noted after some of the m...

Page 38: ...ariable resistor 100k located on the CRT PCB is used for this adjustment procedure It may only be touched by a properlyinsulatingscrewdriver Caution Highvoltage Correct adjustment is achieved when the...