Hameg HM 407A Manual Download Page 29

29

Subject to change without notice

coupling is not set to line/mains trigger coupling. Then the
internal trigger source is deactivated. As the external trigger
signal applied at the

 TRIG. EXT 

socket normally has no relation

to the signal height of the displayed signal, the trigger point
symbol is switched off. The external trigger voltage must have
a minimum amplitude of 0.3Vpp and should not increase above
3Vpp. The input impedance of the

 TRIG. EXT. 

socket is

approx. 1M

 II 15pF.

The maximum input voltage of the input circuit is 100V
(DC+peak AC).

The external trigger voltage may have a completely different
form from the test signal voltage, but must be synchronous
with the test signal. Triggering is even possible in certain limits
with whole number multiples or fractions of the test frequency.

It must be noted that a different phase angle between the
measuring and the triggering signal may cause a display not
coinciding with the slope selection setting.

The trigger coupling selection can also be used in external
triggering mode.

Trigger indicator “TR”

The following description applies to the

 “TR” LED

.

 Please note

item (16) under “Controls and readout”.

An LED on condition indicates that the trigger signal has a
sufficient amplitude and the trigger level control setting is
correct. This is valid with automatic and with normal triggering.
By observing the trigger LED, sensitive trigger level adjustment
is possible when normal triggering is used, particularly at very
low signal frequencies. The indication pulses are of only
100ms duration.

Thus for fast signals the LED appears to glow continuously, for
low repetition rate signals, the LED flashes at the repetition rate
or at a display of several signal periods  not only at the start of
the sweep at the left screen edge, but also at each signal period.

In automatic triggering mode the sweep generator starts
repeatedly without test signal or external trigger voltage. If the
trigger signal frequency decreases the trigger automatic
repetition frequency the sweep generator starts without
awaiting the trigger pulse. This causes an untriggered display
and a flashing trigger LED.

HOLD OFF-time adjustment
(Only in analog mode)

For instrument specific information please note

 DEL.POS. -

HO (27) 

in section “Controls and readout”.

If it is found that a trigger point cannot be found on extremely
complex signals, even after careful adjustment of the trigger
level control, a stable display may often be obtained using the
holdoff control. This facility varies the holdoff time between
two sweep periods approx. up to the ratio 10:1. Pulses or other
signal waveforms appearing during this off period cannot
trigger the time base.

Particularly with burst signals or aperiodic pulse trains of the
same amplitude, the start of the sweep can be delayed until
the optimum or required time.

A very noisy signal or a signal with a higher interfering
frequency is at times displayed double. It is possible that
trigger level adjustment only controls the mutual phase shift,
but not the double display. The stable single display of the

signal, required for evaluation, is easily obtainable by expanding
the hold off time until one signal is displayed.

A double display is possible with certain pulse signals, where
the pulses alternately show a small difference of the peak
amplitudes. Only a very exact trigger level adjustment makes
a single display possible. The use of the holdoff control
simplifies the right adjustment.

After specific use the holdoff control should be reset into its
calibration detent (fully ccw), otherwise the brightness of the
display is reduced drastically. The function is shown in the
following figures.

Fig. 1 shows a case where the holdoff control is in the minimum

position and various different waveforms are overlapped on the
screen, making the signal observation unsuccessful.

Fig. 2 shows a case where only the desired parts of the signal are stably

displayed.

Delay / After Delay Triggering

(Only in analog mode)

The instrument specific information regarding this mode is
part of the section “Controls and readout” paragraph

 DEL.POS.

- HO (27), SEA./DEL. - ON/OFF (29) 

and 

 DEL.TRIG. - VAR.

(30)

.

As mentioned before, triggering starts the time base sweep
and unblanks the beam. After the maximum X deflection to the
right, the beam is blanked and flies back to the (left) start
position. After the hold off period the sweep is started
automatically by the automatic trigger or the next trigger
signal. In normal triggering mode the automatic trigger is
switched off and will only start on receipt of a trigger signal.

As the trigger point is always at the trace start position, trace
expansion in X direction with the aid of the time base is limited
to the display on the left of the trace. Parts of the signal to be
expanded which are displayed near the trace end (right side of
the screen) are lost when the time base speed is increased
(time coefficient reduced).

The delay function delays the trace start by a variable time from
the trigger point. This allows the sweep to begin on any portion
of a signal. The time base speed can then be increased to
expand the display in X direction. With higher expansion rates,
the intensity reduces and within certain limits this can be
compensated by the

 INTENS 

knob setting.

Triggering and time base

Summary of Contents for HM 407A

Page 1: ...in analog mode 29 Auto Set 31 Save Recall 31 Component Tester analog mode 31 General 31 Using the Component Tester 32 Test Procedure 32 Test Pattern Displays 32 Testing Resistors 32 Testing Capacitors...

Page 2: ...mity test made by HAMEG is based on the actual generic and product standards In cases where different limit values are applicable HAMEG applies the severer standard For emission the limits for residen...

Page 3: ...100MHz 0 5div Normal DC 100MHz LED for trigger indication Slope positive or negative Sources CH I or II line ext CH I alternate CH II 0 8div Coupling AC 10Hz 100MHz DC 0 100MHz HF 50kHz 100MHz LF 0 1...

Page 4: ...protective earth contact The protective action must not be negated by the use of an extension cord without a protective conductor Themains lineplugmustbeinsertedbeforeconnections are made to measuring...

Page 5: ...is stated for possible queries this helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain int...

Page 6: ...he readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general electrical engineering alternating voltage data normally refers to effective values rms root mean squ...

Page 7: ...ore the measurement It can lie below or above the horizon tal central line according to whether positive and or negative deviations from the ground potential are to be measured Total value of input vo...

Page 8: ...has little meaning For amplifiers with approximately constant group delay therefore good pulse transmission performance the following numerical relationship between rise time tr in ns and bandwidth B...

Page 9: ...tective conductor connections caused by external equipment connected to the mains line e g signal generators with interference protection capacitors Controls and readout The following description assu...

Page 10: ...a pre or post trigger value PT is not indicated by the readout Pressing and holding the STOR ON button switches over to the digital mode but without changing the channel operating mode CH I CH II DUAL...

Page 11: ...and back to PT0 The values refer to the X axis graticule of the screen display 10 1div The following description assumes that the X magnifier x10 is inactive and the signal display starts on the left...

Page 12: ...V In this case also several signal acquisition scans are required hence it is similar to Refresh operation The signal is averaged over the several acquisitions so that amplitude variations e g noise a...

Page 13: ...r pretrigger and post trigger settings the signal acquisition is not complete when the trigger occurs and will only be terminated later After termination the RES LED extinguishes but the signal displa...

Page 14: ...nob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to analog XY mode this control knob is inactive and the X POS knob must be used for a horizontal po...

Page 15: ...trigger point setting voltage The trigger unit starts the time base when the edge of a trigger signal crosses the trigger point In most Yt modes the trigger point is displayed in the readout by the s...

Page 16: ...nged All controls related to both channels are active if the inputs 31 and 35 are not set to GD 33 37 Whether alternated or chopped channel switching is present depends on the actual time base setting...

Page 17: ...ing in DUAL mode Under these conditions both TRIG CH I and CH II LEDs are lit As alternate triggering requires alternate channel operation alternate channel switching is set automatically A change of...

Page 18: ...of the trace can be set with the DEL POS control in the time base modes SEA SEARCH or DEL DELAY See SEA DEL ON OFF 29 STORAGE MODE ONLY The DEL POS control and the HO LED are disabled since in this m...

Page 19: ...eviously active settings Automatic normal triggering 15 trigger LEVEL 17 trigger slope 15 and trigger coupling 26 will be stored The instrument will be automatically switched to normal triggering NM a...

Page 20: ...as an earth ground symbol instead of the deflection coefficient and the AC or DC symbol The GD setting disables the input signal the AC DC 36 pushbutton and the VOLTS DIV 24 knob In automatic trigger...

Page 21: ...In this mode two signals are normally applied causing an X and a Y deflection The deflection coefficient selected for each channel may be different thus as in DUAL mode the V cursor measurement requi...

Page 22: ...ent the cursor moves fast 44 CAL Pushbutton and concentric socket A squarewave signal of 0 2Vpp 1 is available from the socket for probe adjustment purposes The signal frequency depends on the pushbut...

Page 23: ...eady for use If the AUTO SET function was not used and only a spot appears CAUTION CRT phosphor can be damaged reduce the intensity immediately and check that the XY mode is not selected XY not displa...

Page 24: ...displayed on the CRT screen should have the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator out...

Page 25: ...n the result The following must be noted here Becauseoftheperiodicnatureofthetrigonometricfunctions the calculation should be limited to angles 90 However here is the advantage of the method Due to ph...

Page 26: ...g signal AC voltage are displayable in Yt mode In this mode the signal voltage deflects the beam in vertical direction Y while the time base generator moves the beam from the left to the right of the...

Page 27: ...e previous slope setting for the undelayed time base trigger is stored and still active For further information please note Controls and readout The time base generator can be triggered by a rising or...

Page 28: ...lit 26 The trigger point symbol is inactive in line mains trigger mode as there is no direct amplitude relationship between the trigger voltage and the signal voltage A voltage originating from mains...

Page 29: ...sweep periods approx up to the ratio 10 1 Pulses or other signal waveforms appearing during this off period cannot trigger the time base Particularly with burst signals or aperiodic pulse trains of t...

Page 30: ...tion coefficient must be increased with TIME DIV knob A larger deflection coefficient than in the SEARCH mode cannot be set Example The SEARCH setting selected in figure 2 is 5ms cm The display in DEL...

Page 31: ...veral frequencies e g video signals are present AUTO SET sets the instrument automatically to the following operating conditions The input coupling is not affected while AC or DC is used but if GD was...

Page 32: ...e will tend towards the horizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can be approximately evaluated The determination of actua...

Page 33: ...n circuit tests are possible in many cases However they are not well defined This is caused by a shunt connection of real or complex impedances especially if they are of relatively low impedance at 50...

Page 34: ...rizontal center graticule line by using the Y POS control Then the trigger point symbol may be set 2 division above the 0 Volt position using the LEVEL control If the deflection coefficient is set to...

Page 35: ...commended to switch on the instrument for about 20 minutes prior to the commencement of any check Cathode Ray Tube Brightness and Focus Linearity Raster Distortion Normally the CRT of the instrument h...

Page 36: ...regular intervals the deflection coefficients on all positions of the input attenuators and readjust them as necessary A compensated 2 1 series attenuator is also necessary and this must be matched t...

Page 37: ...For precise determination set the peak of the first marker or cycle peak exactly behind the first vertical graticule line using the X POS control Deviation tendencies can be noted after some of the m...

Page 38: ...ariable resistor 100k located on the CRT PCB is used for this adjustment procedure It may only be touched by a properlyinsulatingscrewdriver Caution Highvoltage Correct adjustment is achieved when the...

Page 39: ...a high level reference potential is not permitted and endangers operator oscilloscope interface and peripheral devices In case of disregard of the safety warnings contained in this manual HAMEG refuse...

Page 40: ...40 Subject to change without notice Front Panel HM407...

Reviews: