38

Subject to change without notice

panel can withdrawn forward. When the chassis is inserted
into the case later on, it should be noticed that the case has to
fit under the flange of the front panel. The same applies for the
rear of the case, on which the rear cover is put.

Caution !
During opening or closing of the case, the instrument
must be disconnected from all power sources for
maintenance work or a change of parts or components.
If a measurement, trouble-shooting, or an adjustment
is unavoidable, this work must be done by a specialist,
who is familiar with the risk involved.

When the instrument is set into operation after the case has been
removed, attention must be paid to the acceleration voltage for
the CRT  approx. -2025V and to the operating voltages for both
final amplifier stages 175V and 146V. Potentials of these voltages
are on the PS-Board, the CRT-PCB and on the main PCB. They are
highly dangerous and therefore precautions must be taken. It
should be noted furthermore that shorts occurring on different
points of the CRT high voltage and unblanking circuitry will
definitely damage some semiconductors. For the same reason it
is very risky to connect capacitors to these points while the
instrument is on.

Capacitors in the instrument may still be charged, even when the
instrument is disconnected from all voltage sources. Normally,
the capacitors are discharged approx. 6 seconds after switching
off. However, with a defective instrument an interruption of the
load is not impossible. Therefore, after switching off, it is
recommended to connect one by one all terminals across 1k

 to

ground (chassis) for a period of 1 second.

Caution

Handling of the CRT needs utmost caution. The glass bulb
must not be allowed  under any circumstances  to come into
contact with hardened tools, nor should it undergo local
superheating (e.g. by soldering iron) or local undercooling (e.g.
by cryogenic-spray). We recommend the wearing of safety
goggles (implosion danger). If any work has been carried out,
the complete instrument (with case closed and

 POWER

button depressed) must undergo a safety test, including a
voltage test with 2200V DC, between accessible parts to both
mains/line supply terminals. This test is dangerous and requires
an adequately trained specialist. The maximum impedance
between each accessible metallic part and the safety earth
connector at the 3-pole power connector must be less than
0.1

.

Operating Voltages

The operating voltages ( +12V, -6V, -2025V) are stabilized by
the switch mode power supply. The +12V supply is further
stabilized and used as a reference voltage for -6V stabilization.
These different operating voltages are fixed voltages, except
the +12V, which can be adjusted. The variation of the fixed
voltages greater than 5% from the nominal value indicates a
fault. Measurements of the high voltage may only be
accomplished by the use of a sufficient highly resistive voltmeter
(>10M

). You must make absolutely sure that the electric

strength of the voltmeter is sufficiently high. It is recommended
to check the ripple and also the interaction from other possible
sources. Excessive values might be very often the reason for
incomprehensible faults.

Maximum and Minimum Brightness

A variable resistor (100k

), located on the CRT PCB, is used

for this adjustment procedure. It may only be touched by a
properly insulating screwdriver (Caution! High voltage!). Correct

adjustment is achieved, when the trace can be blanked in XY
analog mode and, in addition, when the requirement described
in the Test Instructions are met.

Astigmatism control

The ratio of vertical and horizontal sharpness can be adjusted
by the variable resistor of 47k

, located on the CRT PCB. As

a precaution however, the voltage for the vertical deflecting
plates (approx. +71V when the trace is in center position)
should firstly be checked, because this voltage will affect the
astigmatism correction. While the adjustment is being carried
out (with medium brightness and a 1MHz square-wave signal),
the upper horizontal square-wave tops are firstly focussed
with the FOCUS control. Then the sharpness of the vertical
lines are corrected with the 47k

 Astigm. pot. The correction

should be repeated several times in this sequence. The
adjustment is optimized, when the FOCUS knob exclusively
brings no improvement of the sharpness in both directions.

Trigger Threshold

The internal trigger threshold should be in the range 0.3 to
0.5div display height.

Trouble-Shooting the Instrument

For this job, at least an isolating variable mains/line transformer
(protection class II), a signal generator, an adequate precise
multimeter, and, if possible, an oscilloscope are needed. This
last item is required for complex faults, which can be traced by
the display of signal or ripple voltages. As noted before, the
regulated high voltage and the supply voltages for the final
stages are highly dangerous. Therefore it is recommended to
use totally insulated extended probe tips, when trouble-
shooting the instrument. Accidental contact with dangerous
voltage potentials is then unlikely. Of course, these instructions
cannot thoroughly cover all kinds of faults. Some common-
sense will certainly be required, when a complex fault has to
be investigated.

If trouble is suspected, visually inspect the instrument
thoroughly after removal of the case. Look for loose or badly
contacted or discolored components (caused by overheating).
Check to see that all circuit board connections are making good
contact and are not shorting to an adjacent circuit. Especially
inspect the connections between the PCBs. This visual
inspection can lead to success much more quickly than a
systematic fault location using measuring instruments. Prior
to any extensive trouble-shooting, also check the external
power source.

If the instrument fails completely, the first and important step
- after checking the power fuses - will be to measure the
deflecting plate voltages of the CRT. In almost any case, the
faulty section can be located. The sections represent:

1. Vertical deflection.
2. Horizontal deflection.
3. CRT circuit.
4. Power supply.

While the measurement takes place, the position controls of
both deflection devices must be in mid-position. When the
deflection devices are operating properly, the separate voltages
of each plate pair are almost equal then (Y approx. 71V and X
approx 90V). If the separate voltages of a plate pair are very
different, the associated circuit must be faulty. An absent
trace in spite of correct plate voltages means a fault in the CRT
circuit. Missing deflection plate voltages is probably caused by
a defect in the power supply.

Service Instructions

Summary of Contents for HM 407A

Page 1: ...in analog mode 29 Auto Set 31 Save Recall 31 Component Tester analog mode 31 General 31 Using the Component Tester 32 Test Procedure 32 Test Pattern Displays 32 Testing Resistors 32 Testing Capacitors...

Page 2: ...mity test made by HAMEG is based on the actual generic and product standards In cases where different limit values are applicable HAMEG applies the severer standard For emission the limits for residen...

Page 3: ...100MHz 0 5div Normal DC 100MHz LED for trigger indication Slope positive or negative Sources CH I or II line ext CH I alternate CH II 0 8div Coupling AC 10Hz 100MHz DC 0 100MHz HF 50kHz 100MHz LF 0 1...

Page 4: ...protective earth contact The protective action must not be negated by the use of an extension cord without a protective conductor Themains lineplugmustbeinsertedbeforeconnections are made to measuring...

Page 5: ...is stated for possible queries this helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain int...

Page 6: ...he readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general electrical engineering alternating voltage data normally refers to effective values rms root mean squ...

Page 7: ...ore the measurement It can lie below or above the horizon tal central line according to whether positive and or negative deviations from the ground potential are to be measured Total value of input vo...

Page 8: ...has little meaning For amplifiers with approximately constant group delay therefore good pulse transmission performance the following numerical relationship between rise time tr in ns and bandwidth B...

Page 9: ...tective conductor connections caused by external equipment connected to the mains line e g signal generators with interference protection capacitors Controls and readout The following description assu...

Page 10: ...a pre or post trigger value PT is not indicated by the readout Pressing and holding the STOR ON button switches over to the digital mode but without changing the channel operating mode CH I CH II DUAL...

Page 11: ...and back to PT0 The values refer to the X axis graticule of the screen display 10 1div The following description assumes that the X magnifier x10 is inactive and the signal display starts on the left...

Page 12: ...V In this case also several signal acquisition scans are required hence it is similar to Refresh operation The signal is averaged over the several acquisitions so that amplitude variations e g noise a...

Page 13: ...r pretrigger and post trigger settings the signal acquisition is not complete when the trigger occurs and will only be terminated later After termination the RES LED extinguishes but the signal displa...

Page 14: ...nob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to analog XY mode this control knob is inactive and the X POS knob must be used for a horizontal po...

Page 15: ...trigger point setting voltage The trigger unit starts the time base when the edge of a trigger signal crosses the trigger point In most Yt modes the trigger point is displayed in the readout by the s...

Page 16: ...nged All controls related to both channels are active if the inputs 31 and 35 are not set to GD 33 37 Whether alternated or chopped channel switching is present depends on the actual time base setting...

Page 17: ...ing in DUAL mode Under these conditions both TRIG CH I and CH II LEDs are lit As alternate triggering requires alternate channel operation alternate channel switching is set automatically A change of...

Page 18: ...of the trace can be set with the DEL POS control in the time base modes SEA SEARCH or DEL DELAY See SEA DEL ON OFF 29 STORAGE MODE ONLY The DEL POS control and the HO LED are disabled since in this m...

Page 19: ...eviously active settings Automatic normal triggering 15 trigger LEVEL 17 trigger slope 15 and trigger coupling 26 will be stored The instrument will be automatically switched to normal triggering NM a...

Page 20: ...as an earth ground symbol instead of the deflection coefficient and the AC or DC symbol The GD setting disables the input signal the AC DC 36 pushbutton and the VOLTS DIV 24 knob In automatic trigger...

Page 21: ...In this mode two signals are normally applied causing an X and a Y deflection The deflection coefficient selected for each channel may be different thus as in DUAL mode the V cursor measurement requi...

Page 22: ...ent the cursor moves fast 44 CAL Pushbutton and concentric socket A squarewave signal of 0 2Vpp 1 is available from the socket for probe adjustment purposes The signal frequency depends on the pushbut...

Page 23: ...eady for use If the AUTO SET function was not used and only a spot appears CAUTION CRT phosphor can be damaged reduce the intensity immediately and check that the XY mode is not selected XY not displa...

Page 24: ...displayed on the CRT screen should have the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator out...

Page 25: ...n the result The following must be noted here Becauseoftheperiodicnatureofthetrigonometricfunctions the calculation should be limited to angles 90 However here is the advantage of the method Due to ph...

Page 26: ...g signal AC voltage are displayable in Yt mode In this mode the signal voltage deflects the beam in vertical direction Y while the time base generator moves the beam from the left to the right of the...

Page 27: ...e previous slope setting for the undelayed time base trigger is stored and still active For further information please note Controls and readout The time base generator can be triggered by a rising or...

Page 28: ...lit 26 The trigger point symbol is inactive in line mains trigger mode as there is no direct amplitude relationship between the trigger voltage and the signal voltage A voltage originating from mains...

Page 29: ...sweep periods approx up to the ratio 10 1 Pulses or other signal waveforms appearing during this off period cannot trigger the time base Particularly with burst signals or aperiodic pulse trains of t...

Page 30: ...tion coefficient must be increased with TIME DIV knob A larger deflection coefficient than in the SEARCH mode cannot be set Example The SEARCH setting selected in figure 2 is 5ms cm The display in DEL...

Page 31: ...veral frequencies e g video signals are present AUTO SET sets the instrument automatically to the following operating conditions The input coupling is not affected while AC or DC is used but if GD was...

Page 32: ...e will tend towards the horizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can be approximately evaluated The determination of actua...

Page 33: ...n circuit tests are possible in many cases However they are not well defined This is caused by a shunt connection of real or complex impedances especially if they are of relatively low impedance at 50...

Page 34: ...rizontal center graticule line by using the Y POS control Then the trigger point symbol may be set 2 division above the 0 Volt position using the LEVEL control If the deflection coefficient is set to...

Page 35: ...commended to switch on the instrument for about 20 minutes prior to the commencement of any check Cathode Ray Tube Brightness and Focus Linearity Raster Distortion Normally the CRT of the instrument h...

Page 36: ...regular intervals the deflection coefficients on all positions of the input attenuators and readjust them as necessary A compensated 2 1 series attenuator is also necessary and this must be matched t...

Page 37: ...For precise determination set the peak of the first marker or cycle peak exactly behind the first vertical graticule line using the X POS control Deviation tendencies can be noted after some of the m...

Page 38: ...ariable resistor 100k located on the CRT PCB is used for this adjustment procedure It may only be touched by a properlyinsulatingscrewdriver Caution Highvoltage Correct adjustment is achieved when the...

Page 39: ...a high level reference potential is not permitted and endangers operator oscilloscope interface and peripheral devices In case of disregard of the safety warnings contained in this manual HAMEG refuse...