Hameg HM 407A Manual Download Page 35

35

Subject to change without notice

With the start of each recording, the signal voltage at the
input(s) of the analog-to-digital converters is briefly measured
(sampled), converted to an 8-bit value and written to an
address in RAM. The next sampled value is converted in the
same way, but stored at the next RAM address.
The maximum sampling rate is 100MS/s. This yields a sampling
interval of 10ns. Assuming that 10 measurements (samples)
per signal period are sufficient for a sinewave signal, the
maximum frequency to be captured is 10MHz (100ns per
period).

Alias signal display

If, due to the time base setting, the sampling rate is too low,
the display of an alias signal may occur.

For example:
A sinewave signal may be sampled only once per period. If
the sinewave signal frequency is accidentally in phase with
the sampling frequency and each sample is taken at the
positive peak value, a straight line is displayed in this position.

Alias signal display may also occur in the form of an apparent
untriggered waveform display of different frequency from the
true signal. Another aliasing condition is the display of signals
seeming to be amplitude modulated.

The easiest way to recognize alias signals is to switch to analog
mode, where the true waveform is displayed. Transfer from
analog to store mode without changing time base range must
produce the same frequency display.

Operating modes of the vertical amplifiers

In principle, the instrument can operate in digital storage mode
with the same operating modes as in analog mode. Thus, the
following can be displayed:

-

Channel I by itself

-

Channel II by itself

-

Channel I and II simultaneously

-

The sum or difference of both channels

-

XY mode

Storage mode differs from analog operation of the scope in the
following respects:

-

In DUAL mode (for simultaneous display of both channels),
both input signals are also simultaneously captured, since
each channel has its own A/D converter. In contrast to
analog mode therefore, it is not necessary to switch between
chopped and alternating mode.

-

Because of the high display rate (approx. 80Hz) of the screen,
no flicker can occur.

-

The storage time base is also active in XY storage mode.

Test Instructions

General

These Test Instructions are intended as an aid for checking the
most important characteristics of the instrument at regular
intervals without the need for expensive test equipment.
Resulting corrections and readjustments inside the instrument,
indicated by the following tests, are described in the “Service
Manual”. They should only be undertaken by qualified personnel.

A chargeable “Service Manual” is available from

 HAMEG

. The

“Service Manual” describes the complete adjustment of the
instrument in English language and contains the circuit diagrams
as well as the component locations.

As with the First Time Operation instructions, care should be
taken that all variable functions are set to their calibrated
positions. This is the case if

 AUTO SET 

had been pressed. It

is recommended to switch on the instrument for about 20
minutes prior to the commencement of any check.

Cathode-Ray Tube: Brightness and Focus,

Linearity, Raster Distortion

Normally, the CRT of the instrument has very good brightness.
Any reduction of this brightness can only be judged visually.
Decreased brightness may be the result of reduced high
voltage. In this case the sensitivity of the vertical amplifiers is
significantly increased .

The control range for maximum and minimum brightness
(intensity) must be such that the beam just disappears before
reaching the left hand stop of the INTENS. control (particularly
when in XY mode), while with the control at the right hand stop
the focus and the line width are just acceptable.

With maximum intensity the time base fly-back must on no
account be visible. Visible trace fault without input signal:
bright dot on the left side or decreasing brightness from left to
right or shortening of the baseline. (Cause: Incorrect Unblanking
Pulse.) It should be noted that with wide variations in brightness,
refocusing is always necessary. Moreover, with maximum
brightness, no “pumping” of the display must occur. If pumping
does occur, it is normally due to a fault in the regulation
circuitry for the high voltage supply. The presetting pots for the
high voltage circuit, minimum and maximum intensity, are only
accessible inside the instrument.

A certain out-of-focus condition in the edge zone of the screen
must be accepted. It is limited by standards of the CRT
manufacturer. The same is valid for tolerances of the
orthogonality, the undeflected spot position, the non-linearity
and the raster distortion in the marginal zone of the screen in
accordance with international standards (see CRT data book).
These limit values are strictly supervised by

 HAMEG

. The

selection of a cathode-ray tube without any tolerances is
practically impossible.

Astigmatism Check

Check whether the horizontal and vertical sharpness of the
display are equal. This is best seen by displaying a square-wave
signal with the repetition rate of approximately 1MHz. Focus
the horizontal tops of the square-wave signal at normal intensity,
then check the sharpness of the vertical edges. If it is possible
to improve this vertical sharpness by turning the FOCUS
control, then an adjustment of the astigmatism control is
necessary. A potentiometer of 47k Ohm is provided inside the
instrument for the correction of astigmatism. A certain loss of
marginal sharpness of the CRT is unavoidable; this is due to the
manufacturing process of the CRT.

Symmetry and Drift of the Vertical Amplifier

Both of these characteristics are substantially determined by
the input stages of the amplifiers.

The symmetry of both channels and the vertical final amplifier
can be checked by inverting Channel I and II (depress the
corresponding INV pushbutton). The vertical position of the
trace should not change by more than 0.5div. However, a
change of 1div is just permissible. Larger deviations indicate
that changes have occurred in the amplifier.

A further check of the vertical amplifier symmetry is possible
by checking the control range of the

 Y-POS. 

controls. A sine-

Test Instructions

Summary of Contents for HM 407A

Page 1: ...in analog mode 29 Auto Set 31 Save Recall 31 Component Tester analog mode 31 General 31 Using the Component Tester 32 Test Procedure 32 Test Pattern Displays 32 Testing Resistors 32 Testing Capacitors...

Page 2: ...mity test made by HAMEG is based on the actual generic and product standards In cases where different limit values are applicable HAMEG applies the severer standard For emission the limits for residen...

Page 3: ...100MHz 0 5div Normal DC 100MHz LED for trigger indication Slope positive or negative Sources CH I or II line ext CH I alternate CH II 0 8div Coupling AC 10Hz 100MHz DC 0 100MHz HF 50kHz 100MHz LF 0 1...

Page 4: ...protective earth contact The protective action must not be negated by the use of an extension cord without a protective conductor Themains lineplugmustbeinsertedbeforeconnections are made to measuring...

Page 5: ...is stated for possible queries this helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain int...

Page 6: ...he readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general electrical engineering alternating voltage data normally refers to effective values rms root mean squ...

Page 7: ...ore the measurement It can lie below or above the horizon tal central line according to whether positive and or negative deviations from the ground potential are to be measured Total value of input vo...

Page 8: ...has little meaning For amplifiers with approximately constant group delay therefore good pulse transmission performance the following numerical relationship between rise time tr in ns and bandwidth B...

Page 9: ...tective conductor connections caused by external equipment connected to the mains line e g signal generators with interference protection capacitors Controls and readout The following description assu...

Page 10: ...a pre or post trigger value PT is not indicated by the readout Pressing and holding the STOR ON button switches over to the digital mode but without changing the channel operating mode CH I CH II DUAL...

Page 11: ...and back to PT0 The values refer to the X axis graticule of the screen display 10 1div The following description assumes that the X magnifier x10 is inactive and the signal display starts on the left...

Page 12: ...V In this case also several signal acquisition scans are required hence it is similar to Refresh operation The signal is averaged over the several acquisitions so that amplitude variations e g noise a...

Page 13: ...r pretrigger and post trigger settings the signal acquisition is not complete when the trigger occurs and will only be terminated later After termination the RES LED extinguishes but the signal displa...

Page 14: ...nob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to analog XY mode this control knob is inactive and the X POS knob must be used for a horizontal po...

Page 15: ...trigger point setting voltage The trigger unit starts the time base when the edge of a trigger signal crosses the trigger point In most Yt modes the trigger point is displayed in the readout by the s...

Page 16: ...nged All controls related to both channels are active if the inputs 31 and 35 are not set to GD 33 37 Whether alternated or chopped channel switching is present depends on the actual time base setting...

Page 17: ...ing in DUAL mode Under these conditions both TRIG CH I and CH II LEDs are lit As alternate triggering requires alternate channel operation alternate channel switching is set automatically A change of...

Page 18: ...of the trace can be set with the DEL POS control in the time base modes SEA SEARCH or DEL DELAY See SEA DEL ON OFF 29 STORAGE MODE ONLY The DEL POS control and the HO LED are disabled since in this m...

Page 19: ...eviously active settings Automatic normal triggering 15 trigger LEVEL 17 trigger slope 15 and trigger coupling 26 will be stored The instrument will be automatically switched to normal triggering NM a...

Page 20: ...as an earth ground symbol instead of the deflection coefficient and the AC or DC symbol The GD setting disables the input signal the AC DC 36 pushbutton and the VOLTS DIV 24 knob In automatic trigger...

Page 21: ...In this mode two signals are normally applied causing an X and a Y deflection The deflection coefficient selected for each channel may be different thus as in DUAL mode the V cursor measurement requi...

Page 22: ...ent the cursor moves fast 44 CAL Pushbutton and concentric socket A squarewave signal of 0 2Vpp 1 is available from the socket for probe adjustment purposes The signal frequency depends on the pushbut...

Page 23: ...eady for use If the AUTO SET function was not used and only a spot appears CAUTION CRT phosphor can be damaged reduce the intensity immediately and check that the XY mode is not selected XY not displa...

Page 24: ...displayed on the CRT screen should have the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator out...

Page 25: ...n the result The following must be noted here Becauseoftheperiodicnatureofthetrigonometricfunctions the calculation should be limited to angles 90 However here is the advantage of the method Due to ph...

Page 26: ...g signal AC voltage are displayable in Yt mode In this mode the signal voltage deflects the beam in vertical direction Y while the time base generator moves the beam from the left to the right of the...

Page 27: ...e previous slope setting for the undelayed time base trigger is stored and still active For further information please note Controls and readout The time base generator can be triggered by a rising or...

Page 28: ...lit 26 The trigger point symbol is inactive in line mains trigger mode as there is no direct amplitude relationship between the trigger voltage and the signal voltage A voltage originating from mains...

Page 29: ...sweep periods approx up to the ratio 10 1 Pulses or other signal waveforms appearing during this off period cannot trigger the time base Particularly with burst signals or aperiodic pulse trains of t...

Page 30: ...tion coefficient must be increased with TIME DIV knob A larger deflection coefficient than in the SEARCH mode cannot be set Example The SEARCH setting selected in figure 2 is 5ms cm The display in DEL...

Page 31: ...veral frequencies e g video signals are present AUTO SET sets the instrument automatically to the following operating conditions The input coupling is not affected while AC or DC is used but if GD was...

Page 32: ...e will tend towards the horizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can be approximately evaluated The determination of actua...

Page 33: ...n circuit tests are possible in many cases However they are not well defined This is caused by a shunt connection of real or complex impedances especially if they are of relatively low impedance at 50...

Page 34: ...rizontal center graticule line by using the Y POS control Then the trigger point symbol may be set 2 division above the 0 Volt position using the LEVEL control If the deflection coefficient is set to...

Page 35: ...commended to switch on the instrument for about 20 minutes prior to the commencement of any check Cathode Ray Tube Brightness and Focus Linearity Raster Distortion Normally the CRT of the instrument h...

Page 36: ...regular intervals the deflection coefficients on all positions of the input attenuators and readjust them as necessary A compensated 2 1 series attenuator is also necessary and this must be matched t...

Page 37: ...For precise determination set the peak of the first marker or cycle peak exactly behind the first vertical graticule line using the X POS control Deviation tendencies can be noted after some of the m...

Page 38: ...ariable resistor 100k located on the CRT PCB is used for this adjustment procedure It may only be touched by a properlyinsulatingscrewdriver Caution Highvoltage Correct adjustment is achieved when the...

Page 39: ...a high level reference potential is not permitted and endangers operator oscilloscope interface and peripheral devices In case of disregard of the safety warnings contained in this manual HAMEG refuse...

Page 40: ...40 Subject to change without notice Front Panel HM407...

Reviews: