Hameg HM 407A Manual Download Page 24

24

Subject to change without notice

Connect the probes (

Types HZ51, 52, 54, 

or

 HZ36

) to the

 CH

input. Set the deflection coefficient to 5mV/div and the input

coupling to DC. The time deflection coefficient should be set
to 0.2ms/div. All deflection coefficients should be calibrated.
Plug the probe tip into the calibrator output socket.

Approximately 2 complete waveform periods are displayed on
the CRT screen. The compensation trimmer should be adjusted.
The location of the low frequency compensation trimmer can
be found in the probe information sheet. Adjust the trimmer
with the insulated screw driver provided, until the tops of the
square wave signal are exactly parallel to the horizontal graticule
lines (see 1kHz diagram). The signal height should then be 4div
± 0.16div (= 4% (oscilloscope 3% and probe 1%). During this
adjustment, the signal edges will remain invisible.

Adjustment at 1MHz

Probes

 HZ51

,

 52 

and

 54 

can also be HF-compensated. They

incorporate resonance de-emphasing networks (R-trimmer in
conjunction with inductances and capacitors) which permit
probe compensation in the range of the upper frequency limit
of the vertical oscilloscope amplifier. Only this compensative
adjustment ensures optimum utilization of the full bandwidth,
together with constant group delay at the high frequency end,
thereby reducing characteristic transient distortion near the
leading edge (e.g. overshoot, rounding, ringing, holes or
bumps) to an absolute minimum.

Using the probes

 HZ51

,

 52 

and

 54

, the full bandwidth of the

oscilloscope can be utilized without risk of unwanted waveform
distortion.

Prerequisite for this HF compensation is a square wave
generator with fast risetime (typically 4ns), and low output
impedance (approx. 50

), providing 0.2V at a frequency of

approx. 1MHz. The calibrator output of this instrument meets
these requirements when the

 CAL. 

pushbutton is depressed.

Connect the probe to CH.I input. Depress the

 CAL. 

pushbutton

for 1MHz. Operate the oscilloscope as described under 1kHz
but select for 0.2µs/div time deflection coefficient setting.

Insert the probe tip into the output socket. A waveform will be
displayed on the CRT screen, with leading and trailing edges
clearly visible. For the HF-adjustment now to be performed, it
will be necessary to observe the rising edge as well as the
upper left corner of the pulse top. The location of the high
frequency compensation trimmer(s) can also be found in the
probe information sheet. These R-trimmer(s) have to be
adjusted such that the beginning of the pulse is as straight as
possible. Overshoot or excessive rounding are unacceptable.
The adjustment is relatively easy if only one adjusting point is
present. In case of several adjusting points the adjustment is
slightly more difficult, but causes a better result. The rising
edge should be as steep as possible, with a pulse top remaining
as straight and horizontal as possible.

After completion of the HF-adjustment, the signal amplitude
displayed on the CRT screen should have the same value as
during the 1kHz adjustment.

Probes other than those mentioned above, normally have a
larger tip diameter and may not fit into the calibrator output.
Whilst it is not difficult for an experienced operator to build a
suitable adapter, it should be pointed out that most of these
probes have a slower risetime with the effect that the total
bandwidth of scope together with probe may fall far below that
of the oscilloscope. Furthermore, the HF-adjustment feature
is nearly always missing so that waveform distortion can not
be entirely excluded. The adjustment sequence must be
followed in the order described, i.e. first at 1kHz, then at
1MHz. The calibrator frequencies should not be used for time
base calibration. The pulse duty cycle deviates from 1:1 ratio.
Prerequisites for precise and easy probe adjustments, as well
as checks of deflection coefficients, are straight horizontal
pulse tops, calibrated pulse amplitude, and zero-potential at
the pulse base. Frequency and duty cycle are relatively uncritical.
For interpretation of transient response, fast pulse risetimes
and low-impedance generator outputs are of particular
importance.

Providing these essential features, as well as switch-selectable
output-frequencies, the calibrator of the instrument can, under
certain conditions, replace expensive squarewave generators
when testing or compensating wideband-attenuators or -
amplifiers. In such a case, the input of an appropriate circuit will
be connected to the CAL.-output via a suitable probe.

The voltage provided at a high-impedance input (1M

 II 15-

30pF) will correspond to the division ratio of the probe used
(10:1 = 20mVpp output). Suitable probes are HZ51, 52, and 54.

Operating modes of the vertical

amplifiers in Yt mode

The most important controls regarding the operation modes of
the vertical amplifiers are the pushbuttons:

 CH I (21)

,

 DUAL

(22) 

and

 CH II (25)

.

 Their functions are described in the section

“ Controls and readout”.

In most cases oscilloscopes are used to display signals in Yt
mode. Then the signal amplitude deflects the beam in vertical
direction while the time base causes an X deflection (from left
to right) at the same time. Thereafter the beam becomes
blanked and fly back occurs.

The following Yt operation modes are available:
Single channel operation of channel I (Mono CH I).
Single channel operation of channel II (Mono CH II).
Two channel operation of channel I and channel II (DUAL).
Two channel operation of channel I and channel II -displaying
the algebraic result as the sum or difference -

 (ADD)

.

The way the channel switching is determined in DUAL mode
depends on the time base setting and is described in the
section “Controls and readout”.

In

 ADD 

mode the signals of both channels are algebraically

added and displayed as one signal. Whether the resulting
display shows the sum  or difference is dependent on the
phase relationship or the polarity of the signals and on the
invert function.

In ADD mode the following combinations are possible for

In-phase input voltages:

    No invert function active = sum.
    Channel II invert function active = difference.

First Time Operation

Summary of Contents for HM 407A

Page 1: ...in analog mode 29 Auto Set 31 Save Recall 31 Component Tester analog mode 31 General 31 Using the Component Tester 32 Test Procedure 32 Test Pattern Displays 32 Testing Resistors 32 Testing Capacitors...

Page 2: ...mity test made by HAMEG is based on the actual generic and product standards In cases where different limit values are applicable HAMEG applies the severer standard For emission the limits for residen...

Page 3: ...100MHz 0 5div Normal DC 100MHz LED for trigger indication Slope positive or negative Sources CH I or II line ext CH I alternate CH II 0 8div Coupling AC 10Hz 100MHz DC 0 100MHz HF 50kHz 100MHz LF 0 1...

Page 4: ...protective earth contact The protective action must not be negated by the use of an extension cord without a protective conductor Themains lineplugmustbeinsertedbeforeconnections are made to measuring...

Page 5: ...is stated for possible queries this helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain int...

Page 6: ...he readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general electrical engineering alternating voltage data normally refers to effective values rms root mean squ...

Page 7: ...ore the measurement It can lie below or above the horizon tal central line according to whether positive and or negative deviations from the ground potential are to be measured Total value of input vo...

Page 8: ...has little meaning For amplifiers with approximately constant group delay therefore good pulse transmission performance the following numerical relationship between rise time tr in ns and bandwidth B...

Page 9: ...tective conductor connections caused by external equipment connected to the mains line e g signal generators with interference protection capacitors Controls and readout The following description assu...

Page 10: ...a pre or post trigger value PT is not indicated by the readout Pressing and holding the STOR ON button switches over to the digital mode but without changing the channel operating mode CH I CH II DUAL...

Page 11: ...and back to PT0 The values refer to the X axis graticule of the screen display 10 1div The following description assumes that the X magnifier x10 is inactive and the signal display starts on the left...

Page 12: ...V In this case also several signal acquisition scans are required hence it is similar to Refresh operation The signal is averaged over the several acquisitions so that amplitude variations e g noise a...

Page 13: ...r pretrigger and post trigger settings the signal acquisition is not complete when the trigger occurs and will only be terminated later After termination the RES LED extinguishes but the signal displa...

Page 14: ...nob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to analog XY mode this control knob is inactive and the X POS knob must be used for a horizontal po...

Page 15: ...trigger point setting voltage The trigger unit starts the time base when the edge of a trigger signal crosses the trigger point In most Yt modes the trigger point is displayed in the readout by the s...

Page 16: ...nged All controls related to both channels are active if the inputs 31 and 35 are not set to GD 33 37 Whether alternated or chopped channel switching is present depends on the actual time base setting...

Page 17: ...ing in DUAL mode Under these conditions both TRIG CH I and CH II LEDs are lit As alternate triggering requires alternate channel operation alternate channel switching is set automatically A change of...

Page 18: ...of the trace can be set with the DEL POS control in the time base modes SEA SEARCH or DEL DELAY See SEA DEL ON OFF 29 STORAGE MODE ONLY The DEL POS control and the HO LED are disabled since in this m...

Page 19: ...eviously active settings Automatic normal triggering 15 trigger LEVEL 17 trigger slope 15 and trigger coupling 26 will be stored The instrument will be automatically switched to normal triggering NM a...

Page 20: ...as an earth ground symbol instead of the deflection coefficient and the AC or DC symbol The GD setting disables the input signal the AC DC 36 pushbutton and the VOLTS DIV 24 knob In automatic trigger...

Page 21: ...In this mode two signals are normally applied causing an X and a Y deflection The deflection coefficient selected for each channel may be different thus as in DUAL mode the V cursor measurement requi...

Page 22: ...ent the cursor moves fast 44 CAL Pushbutton and concentric socket A squarewave signal of 0 2Vpp 1 is available from the socket for probe adjustment purposes The signal frequency depends on the pushbut...

Page 23: ...eady for use If the AUTO SET function was not used and only a spot appears CAUTION CRT phosphor can be damaged reduce the intensity immediately and check that the XY mode is not selected XY not displa...

Page 24: ...displayed on the CRT screen should have the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator out...

Page 25: ...n the result The following must be noted here Becauseoftheperiodicnatureofthetrigonometricfunctions the calculation should be limited to angles 90 However here is the advantage of the method Due to ph...

Page 26: ...g signal AC voltage are displayable in Yt mode In this mode the signal voltage deflects the beam in vertical direction Y while the time base generator moves the beam from the left to the right of the...

Page 27: ...e previous slope setting for the undelayed time base trigger is stored and still active For further information please note Controls and readout The time base generator can be triggered by a rising or...

Page 28: ...lit 26 The trigger point symbol is inactive in line mains trigger mode as there is no direct amplitude relationship between the trigger voltage and the signal voltage A voltage originating from mains...

Page 29: ...sweep periods approx up to the ratio 10 1 Pulses or other signal waveforms appearing during this off period cannot trigger the time base Particularly with burst signals or aperiodic pulse trains of t...

Page 30: ...tion coefficient must be increased with TIME DIV knob A larger deflection coefficient than in the SEARCH mode cannot be set Example The SEARCH setting selected in figure 2 is 5ms cm The display in DEL...

Page 31: ...veral frequencies e g video signals are present AUTO SET sets the instrument automatically to the following operating conditions The input coupling is not affected while AC or DC is used but if GD was...

Page 32: ...e will tend towards the horizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can be approximately evaluated The determination of actua...

Page 33: ...n circuit tests are possible in many cases However they are not well defined This is caused by a shunt connection of real or complex impedances especially if they are of relatively low impedance at 50...

Page 34: ...rizontal center graticule line by using the Y POS control Then the trigger point symbol may be set 2 division above the 0 Volt position using the LEVEL control If the deflection coefficient is set to...

Page 35: ...commended to switch on the instrument for about 20 minutes prior to the commencement of any check Cathode Ray Tube Brightness and Focus Linearity Raster Distortion Normally the CRT of the instrument h...

Page 36: ...regular intervals the deflection coefficients on all positions of the input attenuators and readjust them as necessary A compensated 2 1 series attenuator is also necessary and this must be matched t...

Page 37: ...For precise determination set the peak of the first marker or cycle peak exactly behind the first vertical graticule line using the X POS control Deviation tendencies can be noted after some of the m...

Page 38: ...ariable resistor 100k located on the CRT PCB is used for this adjustment procedure It may only be touched by a properlyinsulatingscrewdriver Caution Highvoltage Correct adjustment is achieved when the...

Page 39: ...a high level reference potential is not permitted and endangers operator oscilloscope interface and peripheral devices In case of disregard of the safety warnings contained in this manual HAMEG refuse...

Page 40: ...40 Subject to change without notice Front Panel HM407...

Reviews: