Hameg HM 407A Manual Download Page 27

27

Subject to change without notice

voltage point on the trigger signal (trigger point), by the
adjustment of the trigger level control. The control range
depends on the peak to peak value of the signal. This trigger
mode is therefore called Automatic Peak (Value)- Triggering.
Operation of the scope needs only correct amplitude and time
base settings, for a constantly visible trace. Automatic mode
is recommended for all uncomplicated measuring tasks.
However, automatic triggering is also the appropriate operation
mode for the “entry” into difficult measuring problems, e.g.
when the test signal is unknown relating to amplitude, frequency
or shape. Presetting of all parameters is now possible with
automatic triggering; the change to normal triggering can
follow thereafter.

The automatic triggering works above 20Hz. The failure of
automatic triggering at frequencies below 20Hz is abrupt.
However, it is not signified by the trigger indicator LED this is still
blinking. Break down of triggering is best recognizable at the left
screen edge (the start of the trace in differing display height).

The automatic peak (value) triggering operates over all variations
or fluctuations of the test signal above 20Hz. However, if the
pulse duty factor of a square-wave signal exceeds a ratio of
100:1, switching over to normal triggering will be necessary.
Automatic triggering is practicable with internal and external
trigger voltage.

Normal Triggering

Information specific to the instrument are given in the articles

NM - AT (15)

,

 LEVEL (17) 

and

 TRIG. MODE (26) 

in the section

”Controls and readout”. The time fine adjustment

  (VAR.) 

and

the hold off time setting assist in triggering under specially
difficult signal conditions. The following description refers to
the Yt (time base) analog mode.

With normal triggering, the sweep can be started by AC signals
within the frequency range defined by the trigger coupling
setting.

In the absence of an adequate trigger signal or when the trigger
controls (particularly the trigger

 LEVEL 

control) are misadjusted,

no trace is visible, i.e. the screen completely blanked.

When using the internal normal triggering mode, it is possible
to trigger at any amplitude point of a signal edge, even with
very complex signal shapes, by adjusting the trigger

 LEVEL

control. If the signal applied at the Y input is used for triggering
(internal trigger source), its adjusting range is directly dependent
on the display height, which should be at least 0.5div. If it is
smaller than 1div, the trigger

 LEVEL 

adjustment needs to be

operated with a sensitive touch. In the external normal triggering
mode, the same applies to approx. 0.3Vpp external trigger
voltage amplitude.

Other measures for triggering of very complex signals are the
use of the time base variable control and

 HOLD OFF 

time

control, hereinafter mentioned.

(Slope)

Please note item

 (15) 

in section “Controls and readout” for

instrument specific information.

The actual slope setting is displayed in the readout. The setting
is not changed by the

 AUTOSET 

function. The slope setting

can be changed in delay mode for the delay time base trigger
unit if the delay trigger function is active. The previous slope
setting for the undelayed time base trigger is stored and still
active.

 For further information please note “Controls and

readout”.

The time base generator can be triggered by a rising or falling
edge of the test signal. Whether the rising or the falling edge
is used for triggering, depends on the slope direction setting.
This is valid with automatic and normal triggering. The positive
slope direction means an edge going from a negative potential
and rising to a positive potential. This has nothing to do with
zero or ground potential and absolute voltage values. The
positive slope may also lie in a negative part of a signal.
However the trigger point may be varied within certain limits
on the chosen edge using the LEVEL control. The slope
direction is always related to the input signal and the non
inverted display.

Trigger coupling

Instrument specific information regarding this item can be
noted in the “Data Sheet”.

 The coupling setting

 ( TRIG. MODE

(26)) 

and indication are described under “Controls and readout”.

As the automatic triggering does not work below 20Hz, normal
triggering should be used in

 DC 

and

 LF 

trigger coupling mode.

The coupling mode and accordingly the frequency range of the
trigger signal should meet the signal requirements.

AC:

This is the most frequently used trigger mode. The
trigger threshold increases below and above the
frequency limits mentioned in the data sheet. This filter
cuts off both the DC content of the trigger signal and the
lowest frequency range.

DC:

In this coupling mode the trigger signal is coupled
galvanically to the trigger unit. Therefore there is no  low
frequency limit.

DC triggering is recommended if the signal is to be
triggered with quite slow processes or if pulse signals
with constantly changing pulse duty factors have to be
displayed.

HF:

In this coupling mode the transmission range equals a
high pass filter. It cuts off the DC content of the trigger
signal and the lower frequency range.

LF:

LF trigger coupling has a low pass filter characteristic. As
in DC trigger coupling, there is no low limit  for the pass
frequency range (galvanic  coupling) in connection with
normal triggering. The trigger signal is coupled through
a capacitor in the automatic (peak value) triggering and
LF trigger coupling. This results in a low frequency limit
which , however, does not disturb as the limit is below
the repetition frequency of the trigger automatic.

The LF trigger coupling is often more suitable for low
frequency signals than DC trigger coupling because the
noise components of the trigger signals are strongly
suppressed. This avoids or reduces , under borderline
conditions, jitter or double traces especially with very
low signal voltages. The trigger threshold raises
continuously above the pass band.

TV-L:

The built-in active TV-Sync-Separator provides the

separation of line sync pulses from the video signal. Even
distorted video signals are triggered and displayed in a
stable manner. This mode is described under paragraph
“Triggering of video signals”.

TV-F:

The built-in active TV-Sync-Separator also provides the

separation of frame sync pulses from the video signal.
Even distorted video signals are triggered and displayed
in a stable manner. This mode is described under
paragraph “Triggering of video signals”.

Triggering and time base

Summary of Contents for HM 407A

Page 1: ...in analog mode 29 Auto Set 31 Save Recall 31 Component Tester analog mode 31 General 31 Using the Component Tester 32 Test Procedure 32 Test Pattern Displays 32 Testing Resistors 32 Testing Capacitors...

Page 2: ...mity test made by HAMEG is based on the actual generic and product standards In cases where different limit values are applicable HAMEG applies the severer standard For emission the limits for residen...

Page 3: ...100MHz 0 5div Normal DC 100MHz LED for trigger indication Slope positive or negative Sources CH I or II line ext CH I alternate CH II 0 8div Coupling AC 10Hz 100MHz DC 0 100MHz HF 50kHz 100MHz LF 0 1...

Page 4: ...protective earth contact The protective action must not be negated by the use of an extension cord without a protective conductor Themains lineplugmustbeinsertedbeforeconnections are made to measuring...

Page 5: ...is stated for possible queries this helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain int...

Page 6: ...he readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general electrical engineering alternating voltage data normally refers to effective values rms root mean squ...

Page 7: ...ore the measurement It can lie below or above the horizon tal central line according to whether positive and or negative deviations from the ground potential are to be measured Total value of input vo...

Page 8: ...has little meaning For amplifiers with approximately constant group delay therefore good pulse transmission performance the following numerical relationship between rise time tr in ns and bandwidth B...

Page 9: ...tective conductor connections caused by external equipment connected to the mains line e g signal generators with interference protection capacitors Controls and readout The following description assu...

Page 10: ...a pre or post trigger value PT is not indicated by the readout Pressing and holding the STOR ON button switches over to the digital mode but without changing the channel operating mode CH I CH II DUAL...

Page 11: ...and back to PT0 The values refer to the X axis graticule of the screen display 10 1div The following description assumes that the X magnifier x10 is inactive and the signal display starts on the left...

Page 12: ...V In this case also several signal acquisition scans are required hence it is similar to Refresh operation The signal is averaged over the several acquisitions so that amplitude variations e g noise a...

Page 13: ...r pretrigger and post trigger settings the signal acquisition is not complete when the trigger occurs and will only be terminated later After termination the RES LED extinguishes but the signal displa...

Page 14: ...nob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to analog XY mode this control knob is inactive and the X POS knob must be used for a horizontal po...

Page 15: ...trigger point setting voltage The trigger unit starts the time base when the edge of a trigger signal crosses the trigger point In most Yt modes the trigger point is displayed in the readout by the s...

Page 16: ...nged All controls related to both channels are active if the inputs 31 and 35 are not set to GD 33 37 Whether alternated or chopped channel switching is present depends on the actual time base setting...

Page 17: ...ing in DUAL mode Under these conditions both TRIG CH I and CH II LEDs are lit As alternate triggering requires alternate channel operation alternate channel switching is set automatically A change of...

Page 18: ...of the trace can be set with the DEL POS control in the time base modes SEA SEARCH or DEL DELAY See SEA DEL ON OFF 29 STORAGE MODE ONLY The DEL POS control and the HO LED are disabled since in this m...

Page 19: ...eviously active settings Automatic normal triggering 15 trigger LEVEL 17 trigger slope 15 and trigger coupling 26 will be stored The instrument will be automatically switched to normal triggering NM a...

Page 20: ...as an earth ground symbol instead of the deflection coefficient and the AC or DC symbol The GD setting disables the input signal the AC DC 36 pushbutton and the VOLTS DIV 24 knob In automatic trigger...

Page 21: ...In this mode two signals are normally applied causing an X and a Y deflection The deflection coefficient selected for each channel may be different thus as in DUAL mode the V cursor measurement requi...

Page 22: ...ent the cursor moves fast 44 CAL Pushbutton and concentric socket A squarewave signal of 0 2Vpp 1 is available from the socket for probe adjustment purposes The signal frequency depends on the pushbut...

Page 23: ...eady for use If the AUTO SET function was not used and only a spot appears CAUTION CRT phosphor can be damaged reduce the intensity immediately and check that the XY mode is not selected XY not displa...

Page 24: ...displayed on the CRT screen should have the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator out...

Page 25: ...n the result The following must be noted here Becauseoftheperiodicnatureofthetrigonometricfunctions the calculation should be limited to angles 90 However here is the advantage of the method Due to ph...

Page 26: ...g signal AC voltage are displayable in Yt mode In this mode the signal voltage deflects the beam in vertical direction Y while the time base generator moves the beam from the left to the right of the...

Page 27: ...e previous slope setting for the undelayed time base trigger is stored and still active For further information please note Controls and readout The time base generator can be triggered by a rising or...

Page 28: ...lit 26 The trigger point symbol is inactive in line mains trigger mode as there is no direct amplitude relationship between the trigger voltage and the signal voltage A voltage originating from mains...

Page 29: ...sweep periods approx up to the ratio 10 1 Pulses or other signal waveforms appearing during this off period cannot trigger the time base Particularly with burst signals or aperiodic pulse trains of t...

Page 30: ...tion coefficient must be increased with TIME DIV knob A larger deflection coefficient than in the SEARCH mode cannot be set Example The SEARCH setting selected in figure 2 is 5ms cm The display in DEL...

Page 31: ...veral frequencies e g video signals are present AUTO SET sets the instrument automatically to the following operating conditions The input coupling is not affected while AC or DC is used but if GD was...

Page 32: ...e will tend towards the horizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can be approximately evaluated The determination of actua...

Page 33: ...n circuit tests are possible in many cases However they are not well defined This is caused by a shunt connection of real or complex impedances especially if they are of relatively low impedance at 50...

Page 34: ...rizontal center graticule line by using the Y POS control Then the trigger point symbol may be set 2 division above the 0 Volt position using the LEVEL control If the deflection coefficient is set to...

Page 35: ...commended to switch on the instrument for about 20 minutes prior to the commencement of any check Cathode Ray Tube Brightness and Focus Linearity Raster Distortion Normally the CRT of the instrument h...

Page 36: ...regular intervals the deflection coefficients on all positions of the input attenuators and readjust them as necessary A compensated 2 1 series attenuator is also necessary and this must be matched t...

Page 37: ...For precise determination set the peak of the first marker or cycle peak exactly behind the first vertical graticule line using the X POS control Deviation tendencies can be noted after some of the m...

Page 38: ...ariable resistor 100k located on the CRT PCB is used for this adjustment procedure It may only be touched by a properlyinsulatingscrewdriver Caution Highvoltage Correct adjustment is achieved when the...

Page 39: ...a high level reference potential is not permitted and endangers operator oscilloscope interface and peripheral devices In case of disregard of the safety warnings contained in this manual HAMEG refuse...

Page 40: ...40 Subject to change without notice Front Panel HM407...

Reviews: