LCR-6000 Series User Manual
87
Figure 9-1
Basic Measurement Procedure
Start
Setup measurement
conditions:
[FUNC]
[LEVEL]
[FREQ]
Connect the test fixture
CORRECTIONS
OPEN TEST
SHORT TEST
Connect DUT
Perform measurement
END
9.2
Example
This paragraph describes a practical example of
measuring a ceramic capacitor.
The basic procedure flow to perform this measurement is
the same as the Basic Measurement Procedure described
previously. In this example, a ceramic capacitor is
measured under the following conditions.
Sample (DUT) Ceramic capacitor Measurement
Conditions:
Function: Cs-D
Test Frequency: 1 kHz
Test Signal Level: 1V