LCR-6000 Series User Manual
37
4.2 [OPEN/SHORT] Page
When you press the [Measure] key and the [OPEN
SHORT] soft-key, the [ OPEN SHORT ] page appears.
In this page, the OPEN/SHORT/SPOT correction for
correcting the stray admittance and residual impedances
can be performed.
The OPEN and SHORT function performs open or
short correction over a range of fixed trimming points.
The correction for all other frequencies are interpolated
from those fixed trimming points. Please refer to the 12.1
Specification page to see what the trimming frequency
points are for each LCR-6000 model.
The SPOT function will perform open, or short
correction at a user-defined “spot” frequency.
In the [ OPEN SHORT] page, you can configure each of
the following controls with the cursor placed in the
corresponding field:
Open correction [OPEN]
Short correction [SHORT]
SPOT correction [SPOT]
Figure 4-2
[OPEN/SHORT] Page