Overview
18
2.4 Main Functions
2.4.1 Correction Function
OPEN/SHORT correction:
Eliminates measurement errors due to stray parasitic
impedance in the test fixtures.
2.4.2 Comparator Function (Sorting)
Bin sort
The primary parameter can be sorted into BIN1-BIN9,
AUX, OUT and HI/IN/LO for each of the primary
measurement parameters.
The sequential mode or tolerance mode can be selected as
the sorting mode.
Limit Setup
Absolute value, deviation value, and % deviation value
can be used for setup.
BIN count
Countable from 0 to 999999
2.4.3 List Sweep
Points
There is a maximum of 10 points.
Sweep parameters
Test frequency, test voltage, test current.
Comparator function of list sweep
The comparator function enables you to set one pair of
lower and upper limits for each measurement point.
You can select from:
Judge with the first sweep parameter/Judge with the
second parameter/Not used for each pair of limits.