Measure Key
40
Step 4. Press the [MEAS SHORT] or [DCR SHORT]
soft-key, a dialog message, “Short-circuit the
test terminals”, will appear.
Step 5. Connect the test fixture to the BNC terminals
and short-circuit the test terminals.
Step 6. Press the [OK] soft-key. The LCR-6000
measures the short impedances (inductances
and resistance) at the all test frequency points.
During the measurement, an “LCR SHORT
measurement in progress” dialog message is
shown on the display.
When the measurement has finished,
“Correction finished” is displayed. During the
measurement, you can press the [ABORT]
soft-key to abort the SHORT correction.
Step 7. After the SHORT measurement has finished,
the meter will enable SHORT correction
automatically. You can, however, still choose
to manually choose to enable or disable
SHORT correction by depressing the ON or
OFF soft-keys.
4.2.3 SPOT Correction
The spot correction function involves performing
open/short correction at user-specified frequency points.
You can specify 1 frequency point.
To specify frequency points and perform open
correction
Step 1. Press the [Measure] key.
Step 3. Press the [OPEN SHORT] soft-key.
Step 3. Use the cursor key to select the [SPOT] field.
Step 4. Enter the frequency using the numeric entry
keys.
Soft-key
Function
ON
Enable this point.
OFF
Disables this point.
CURRENT FREQ Use current frequency
MEAS OPEN
Starts open correction.
MEAS SHORT
Starts short correction