LCR-6000 Series User Manual
39
During the measurement, an “LCR OPEN
measurement in progress” dialog message is
shown on the display.
When the measurement has finished,
“Correction finished” is displayed. During the
measurement, you can press the [ABORT]
soft-key to abort the OPEN correction.
Step 7. After the OPEN measurement has finished,
the meter will enable OPEN correction
automatically. You can, however, still choose
to manually choose to enable or disable OPEN
correction by depressing the ON or OFF soft-
keys.
4.2.2 Short Correction [SHORT]
The short correction feature of the LCR-6000 Series
compensates for any residual impedances (R,X) that may
exist outside of calibration, such as the impedance of the
cables and the DUT connection points.
See Figure 4-4 Residual Impedances.
Figure 4-4
Residual Impedances
To perform short correction
Step 1. Press the [Measure] key.
Step 2. Press the [OPEN SHORT] soft-key.
Step 3. Use the cursor key to select the [SHORT] field.
Soft-key
Function
ON
Enables short correction.
OFF
Disables short correction.
MEAS SHORT
Starts short correction (AC).
DCR SHORT
Starts short correction (DC).