Measure Key
28
Table 4-5
Effective measurement range for the impedance range
when in HOLD state.
Range
No.
Impedance
range
Effective measurement range
8
10Ω
0 - 10Ω
7
30Ω
10Ω-100Ω
6
100Ω
100Ω-316Ω
5
300Ω
316Ω-1kΩ
4
1kΩ
1kΩ-3.16kΩ
3
3kΩ
3.16kΩ-10kΩ
2
10kΩ
10kΩ-31.6kΩ
1
30kΩ
31.6kΩ-100kΩ (
)
0
100kΩ
100kΩ-
(This range is available only
when the test frequency < 20kHz.)
Table 4-6
DCR FUNC effective measurement range for the
impedance range when in HOLD state.
When the measured value exceeds 150% of the full
scale for the selected range, “OVERLOAD” will be
displayed.
Range
No.
Impedance
range
Effective measurement
range
Overload
8
1Ω
0
Ω
– 3.1
Ω
4.65
Ω
7
10Ω
2.8
Ω
– 99
Ω
148.5
Ω
6
100Ω
90
Ω
– 312
Ω
468
Ω
5
300Ω
280
Ω
– 990
Ω
1.485k
Ω
4
1kΩ
900
Ω
– 3.1k
Ω
4.65k
Ω
3
3kΩ
2.8k
Ω
– 9.9k
Ω
14.85k
Ω
2
10kΩ
9k
Ω
– 31k
Ω
46.5k
Ω
1
30kΩ
28k
Ω
– 99k
Ω
148.5k
Ω
0
100kΩ
90k
Ω
– 312k
Ω
100M
Ω
DCR accuracy: ±0.05% with effective measurement.
range