LCR-6000 Series User Manual
137
Figure 13-2
Table for basic accuracy correction factor A
r
10m
20m
50m
5m
100m
200m
500m
1
2
3
1
2
5
10
20
50
100
1.2
150m
Ar
[Vrms]
Table 13-1
Impedance correction factors
Measurement
speed
Measuring frequency K
a
K
b
Slow
Medium
f
m
100Hz
)
100
1
)(
200
1
)(
10
1
(
3
m
s
m
f
V
Z
)
100
1
)(
70
1
)(
10
1
(
9
m
s
m
f
V
Z
100Hz
f
m
≤100kHz
)
200
1
)(
10
1
(
3
s
m
V
Z
)
70
1
)(
10
1
(
9
s
m
V
Z
f
m
100kHz
)
200
2
)(
10
1
(
3
s
m
V
Z
)
70
1
)(
10
3
(
9
s
m
V
Z
Fast
f
m
100Hz
)
100
1
)(
400
1
)(
10
5
.
2
(
3
m
s
m
f
V
Z
)
100
1
)(
100
1
)(
10
2
(
9
m
s
m
f
V
Z
100Hz
f
m
≤100kHz
)
400
1
)(
10
5
.
2
(
3
s
m
V
Z
)
100
1
)(
10
2
(
9
s
m
V
Z
f
m
100kHz
)
400
2
)(
10
5
.
2
(
3
s
m
V
Z
)
100
1
)(
10
6
(
9
s
m
V
Z
f
m
: Measuring frequency [Hz]
Z
m
: The impedance of the DUT [Ω]
V
s
: The amplitude of the measuring signal [mVrms]
When the impedance is less than 500Ω then use K
a
; K
b
can be ignored.
When the impedance is larger than 500Ω then use K
b
; K
a
can be ignored.