LCR-6000 Series User Manual
29
Procedure for setting the impedance range [RANGE]
Step 1. Press the [Measure] key.
Step 2. Use the cursor key to select the [RANGE]
field.
Step 3. Use the soft-keys to select the impedance
range mode or impedance range.
Soft-key
Function
AUTO RANGE
Sets to AUTO range
HOLD RANGE
Sets to HOLD range
NOMINAL RANGE
Sets to NOMINAL range
INCR +
Increments the impedance
range in the HOLD mode
DECR -
Decrements the impedance
range in the HOLD mode
4.1.3 Test Frequency [FREQ]
LCR-6300: 10Hz
~
300kHz
LCR-6200: 10Hz
~
200kHz
LCR-6100: 10Hz
~
100kHz
LCR-6020: 10Hz
~
20kHz
LCR-6002: 10Hz
~
2kHz
Table 4-7
Frequency range and resolution
Frequency range(F)
Resolution
10.00Hz
≦
F
≦
99.99Hz
0.01Hz
100.0Hz
≦
F
≦
999.9Hz
0.1Hz
1.000kHz
≦
F
≦
9.999kHz
1Hz
10.00kHz
≦
F
≦
99.99kHz
10Hz
100.0kHz
≦
F
≦
300.0kHz
100Hz
Frequency Accuracy: ±0.01% with 4-digit resolution