Setup Key
54
5.1.1 Source Output Impedance [SRC RES]
The Source output impedance can be set to 30Ω, 50Ω or
100Ω.
If you use LCR-6000 Series to test a smaller inductor,
please use 30Ω.
If you need to compare test results with Keysight
E4980A, select 100Ω.
Procedure for setting source output impedance
Step 1. Press the [Setup] key.
Step 2. Use the cursor key to select the [SRC RES]
field.
Step 3. Use the soft-keys to set the source output
impedance.
Soft-key
Function
30Ω
30Ω
50Ω
50Ω
100Ω
100Ω
5.1.2 Averaging Factor [AVG]
The averaging feature allows you to obtain moving
average values of successive measurement results.
You can specify the averaging factor from integer 1 to
integer 256.
To set up the averaging factor
Step 1. Press the [Setup] key.
Step 2. Use the cursor key to select [AVG] field.
Step 3. Use the soft-keys or number keys to enter
averaging factor.
Soft-key
Function
INCR +
Increments the averaging factor in
steps of 1, 2, 4, 8, 16, 32,
64, 128 and 256.
DECR -
Decrements the averaging factor in
steps of 1, 2, 4, 8, 16, 32,
64, 128 and 256.