LCR-6000 Series User Manual
49
Soft-key
Function
SAVE & STOP Stops the recording and saves the
recorded readings in the internal
buffer to an external USB flash drive.
The internal buffer will be cleared
after the save operation.
If an external USB flash drive is not
plugged in, the meter will not stop the
ongoing recording.
CLEAR &
STOP
Stop recording and clears the internal
buffer.
Step 5. When the internal buffer is full, “!FULL” will
be added after the number of recorded
counts(in the [LOG] status field). The
following two operations are available in this
situation:
Soft-key
Function
SAVE TO USB Saves the recorded readings in the
buffer to an external USB flash drive.
The internal buffer is cleared after this
operation.
CLEAR
BUFFER
Clears the internal buffer.
Before saving the recorded readings in the internal buffer
to the external USB flash drive, remember to plug the
external USB drive into the USB port on the front panel.
The recorded readings will be saved in a subdirectory
named LIST DATA, whose parent directory will have
the same name as the LCR-6000 model being used. See
the example in below:
Example: F:\LCR6300\LIST DATE\LIST0000.CSV
Up to 9999 recording files can be saved, with their file
names from 0001.csv~9999.csv.
The size of the internal buffer can be adjusted in
SYSTEM→SYSTEM CONFIG→DATA BUFFER