LCR-6000 Series User Manual
43
4.3.2 Test Mode [MODE]
The [LIST MEAS] page will execute a list of up to 10
swept frequency or amplitude tests.
When [MODE] is set to SEQ and [TRIG] is set to
MAN, the [LIST MEAS] function will automatically
execute each test step on the list in sequence until the
last step is executed. The meter will then wait for the
trigger button to be depressed before repeating the
process.
When [MODE] is set to STEP and [TRIG] is set to
MAN, the [LIST MEAS] function will automatically
execute the first test step. The meter will then wait for
the trigger button to be depressed before executing the
next step. This process is repeated for each step on the
list.
Soft-key
Function
SEQ
Sequence measure all step by one
trigger.
STEP
Measure one step by one trigger.
There are 4 available trigger modes in the [LIST MEAS]
page.
Trigger Mode
Function
INT
Trigger by the meter’s internal trigger
source.
MAN
Trigger by pressing the trigger
button.
EXT
Trigger by sending a signal to the
trigger pin on the handler interface.
BUS
Trigger by the RS-232 port.