Measure Key
32
4.1.5 Test Signal Voltage/Current
Level [LEVEL]
The LCR-6000 Series’ test signal voltage/current level
can be set as the effective value (RMS value) of a sine
wave of the test frequency from the unit’s internal
oscillator.
When a constant voltage or current level measurement is
performed, an asterisk mark (*) appears at the end of the
LEVEL display.
Test signal level setting procedure
Step 1. Press the [Measure] key
Step 2. Use the cursor key to select [LEVEL] field
Step 3. Use the soft-keys or numeric entry keys to
enter the test signal voltage/current level.
Soft-key
Function
INCR +
Refer to Table 4-13
DECR -
ALC ON
Auto Level Control ON
ALC OFF
Auto Level Control OFF
Table 4-13
Test voltage/current level that can be selected with
INCR+/DECR
‐
INCR+/DECR-
10.00mV
100.0mV
300.0mV
500.0mV
1.00V
1.50V
2.00V
INCR+/DECR-
100.0uA 500.0uA 1.00mA 5.00mA 10.00mA 20.00mA
Voltage range
Resolution
10.00mV
≦
LEVEL
≦
99.99mV
0.01mV
100.0mV
≦
LEVEL
≦
999.9mV
0.1mV
1.00V
≦
LEVEL
≦
2.00V
0.01V
Current range
Resolution
100.0uA
≦
LEVEL
≦
999.9uA
0.1uA
1.00mA
≦
LEVEL
≦
20.00mA
0.01mA