LCR-6000 Series User Manual
105
11.8 LEVel Subsystem
The Level subsystem sets the oscillator’s output voltage/current level and
source output impedance.
Figure 11-5
LEVel Subsystem Command Tree
11.8.1 LEVel:VOLTage (=VOLTage[:LEVel])
The LEVel:VOLTage or VOLTage[:LEVel] command sets the oscillator’s
output voltage level.
Command
Syntax
LEVel:VOLTage {<value>,MIN,MAX}
or VOLTage:LEVel {<value>,MIN,MAX}
Parameter
Where,
<value> Is the numeric data (NR1, NR2 or NR3).
MIN
Sets to the minimum value of voltage.
MAX Sets to the maximum value of voltage.
Example
SEND> LEV:VOLT 0.3<NL>//Set to 0.3V, V unit
suffix cannot be used.
Query Syntax
LEVel:VOLTage?
or VOLTage:LEVel?
Query Response <NR3>
NR3 floating point
Example
SEND> VOLT?<NL>
RET>
1.000e+00<NL>
Note
The V unit suffix can’t be used.
This command CANNOT be used in LIST MEAS page
and CORRECTION page.