Figure 48: Radar Config - Position Mode.
RADAR CONFIG > SCAN > POSITION OFFSET (ns)
This parameter defines the location of Time Zero displayed on the SIR 30 with respect to the direct
coupling. The SIR 30 will position the signal at the first positive peak of the direct coupling signal at zero
time. This parameter can only be modified if the Position Mode is set to Manual.
Figure 49: Radar Config - Time Zero.
RADAR CONFIG > SCAN > SURFACE %
This parameter defines the total time displayed on the SIR 30 as a percentage of the total time range
selected. In Figure 48 below the Position Mode is set to Auto, The time range has been set to 10 ns. The
Surface % has been set (by default) at 10. The total time displayed is 9 ns. The total time recorded is 10 ns.
Time Zero