IMPORTANT:
There is one DCOM setting on the SIR 30 Host (blue box) that must be set by hand. (It is
not necessary to set this on the Client). The Maintenance tool cannot access it. The setting has been
applied at the factory by GSSI but if it is changed or if SIR 30 software is uninstalled on the SIR 30 Host,
it would need to be set up by hand. To check this setting:
1
Insert the SIR 30 Maintenance tool (USB flash drive into your computer).
2
Run the SIR 30 Maintenance tool (Right-click Run As Administrator).
3
Navigate to the DCOM Tab and click Open DCOM Config.
4
On the Component Service Dialog select: Component Services -> Computers -> My Computer ->
DCOM Config the dialog should look like the figure below:
Figure 165: Component Service Dialog.
5
The User will need to scroll down the list until you find SIR4XSVR. Right-click on it and select
Properties.