Audio alarm specifications
The table below shows the specifications of the audio alarms:
Type of alarm
Repetitions
Frequency of chimes
High
Low
Insistent
high/high/low chime
3,000Hz
2,200Hz
every 2.25 seconds.
Non-insistent
high/low chime
3,000Hz
2,200Hz
every 2.25 seconds.
Back Up
high pitched3,000 Hz
every 1.5 second.
Continuous
high pitched3,000 Hz
without pause.
Enabling special programming functions
In software versions before version 0.71, the Rate Taper, Volume over
Time and Dose-Rate Calculation pump programming modes could
only be enabled or disabled on the Biomedical Menu of Dose-Rate
pumps.
From version 0.71, these parameters appear on the Technician Menu
described on page 4-2, as well as on the Biomedical Menu.
It is recommended that you should place an external label on a pump
to indicate which functions have been enabled.
Enable/disable rate taper
Enabling this parameter allows the user to taper the infusion rate,
either up or down, for the administration of Total Parenteral Nutrition
(TPN), Total Parenteral Admixture (Three-in-One) or other applicable
therapies.
Use the
*
key to toggle this parameter. Thus if you see a screen like
the following:
then the rate taper is enabled and you must press the
*
key to disable
it.
If you press
*
, the screen changes to:
Press the
*
key again to re-enable the Rate Taper mode.
Summary of Contents for 3000/500
Page 1: ...0473 Technical Service Manual Model 3000 500 and Micro 3100 505 Volumetric Infusion Pump...
Page 2: ......
Page 14: ...I n t r o d u c t i o n Chapter 1...
Page 18: ...S p e c i f i c a t i o n Chapter 2...
Page 29: ...O v e r v i e w o f P u m p O p e r a t i o n I n i t i a l C h e c k s Chapter 3...
Page 47: ...V o l u m e t r i c P u m p M e n u s Chapter 4...
Page 79: ...M e c h a n i c a l S y s t e m s Chapter 5...
Page 95: ...E l e c t r o n i c C i r c u i t s Chapter 6...
Page 96: ...M a i n t e n a n c e P r o c e d u r e s T r o u b l e s h o o t i n g Chapter 7...
Page 151: ...I l l u s t r a t e d P a r t s L i s t Chapter 9...
Page 178: ...E l e c t r i c a l D i a g r a m s Chapter 10...
Page 189: ...W a v e f o r m D i a g r a m s Chapter 11...
Page 198: ...R e m o t e M o n i t o r i n g C o n t r o l o f t h e p u m p Chapter 12...
Page 225: ...I n d e x...