About the Technician menu
This menu appears in version 0.71 software (or later). It allows you to
define how the pump behaves by setting values or by disabling or
enabling parameters.
Common parameters on Technician and Biomedical menus:
The common parameters on both the Technician and Biomedical
menus are Rate Taper, Volume Over Time and Dose Rate
Calculation.
Entering the Technician menu
To enter this menu, first switch off the pump if it is switched on.
Then press the
Total Volume
key while simultaneously pressing the
On/Off
key. When you take your fingers off the keys, the pump
switches on and you are asked to enter the Technician menu ID.
Enter the numbers
0002
.
As you enter the digits, asterisks (*) are displayed for security
protection. If you make a mistake, you must switch the pump off to
enter the correct ID. When prompted:
Enter the password
2020
.
The password is always 2020, and cannot be changed.
Using the Technician menu
When you enter the Technician menu, the first parameter is the one
to enable or disable the display of the
Pump Identifier
.
Move down
the list by pressing the
Options
key. Use the
*
key to toggle a
parameter. Thus if you see a screen like the following:
then the feature is disabled and you must press the
*
key to enable it.
You cannot move back through the menu. If you make a mistake, then
you can continue to move through the list by continuing to press the
Options
key and the first item appears again when you reach the
end. You may find it faster to switch off and start again.
To move through the list quickly, and avoid going past the item you
want, you may also find it faster to count the number of times you
press the
Options
key. The table on the next page has been
numbered to help you do this.
Leaving the Technician menu
To exit from the menu, switch off the pump.
Summary of Contents for 3000/500
Page 1: ...0473 Technical Service Manual Model 3000 500 and Micro 3100 505 Volumetric Infusion Pump...
Page 2: ......
Page 14: ...I n t r o d u c t i o n Chapter 1...
Page 18: ...S p e c i f i c a t i o n Chapter 2...
Page 29: ...O v e r v i e w o f P u m p O p e r a t i o n I n i t i a l C h e c k s Chapter 3...
Page 47: ...V o l u m e t r i c P u m p M e n u s Chapter 4...
Page 79: ...M e c h a n i c a l S y s t e m s Chapter 5...
Page 95: ...E l e c t r o n i c C i r c u i t s Chapter 6...
Page 96: ...M a i n t e n a n c e P r o c e d u r e s T r o u b l e s h o o t i n g Chapter 7...
Page 151: ...I l l u s t r a t e d P a r t s L i s t Chapter 9...
Page 178: ...E l e c t r i c a l D i a g r a m s Chapter 10...
Page 189: ...W a v e f o r m D i a g r a m s Chapter 11...
Page 198: ...R e m o t e M o n i t o r i n g C o n t r o l o f t h e p u m p Chapter 12...
Page 225: ...I n d e x...