20. WARNING: The Secondary volume to be infused must match the amount of fluid in the secondary
container. Primary flow resumes when the secondary container is empty. If the volumes do not
correspond, the wrong infusion may be delivered which could cause serious patient injury or death.
21. WARNING: Correct management of battery charging, as described in this documentation is essential to
ensure that the pump can operate on battery for the time specified. Failure to do so may result in
compromised function of the product or patient injury.
22. WARNING: If a backup alarm sounds, the pump should be immediately removed from the patient and sent
to be repaired by a Graseby Medical qualified technician. Failure to do so may cause patient injury or
death.
23. WARNING: Failure to use the power cord retainer means that the pump may be accidentally or erroneously
disconnected from the mains. Although there is a battery backup in case this happens, the battery may not
be charged sufficiently. Consequently, there is a risk of the pump not functioning which could lead to
patient injury or death.
24. WARNING: Do not open the pump housing. Refer all service faults only to qualified technical personnel.
Opening the pump housing may cause electric shock leading to patient or user injury or death.
25. WARNING: When the pump is carrying out an infusion, to ensure that electrical safety is maintained, only
items of equipment that conform to EN60950 are to be connected to the RS232 connector situated at the
back of the pump, otherwise patient safety may be compromised.
26. WARNING: While Graseby Medical Limited have taken all reasonable steps to ensure that the pump
operates correctly while under remote control, it is the responsibility of the person who designs and
implements the controlling device to ensure that the resulting system (pump and controlling device) is fit
for its intended purpose. Failure to do so may result in compromised function of the product, patient injury
or user injury.
27. WARNING: Use only Graseby Medical administration sets with this product. Failure to do so may result in
compromised system accuracy leading to complications resulting in patient injury or death.
28. WARNING: Do not perform these tests while the pump is in use on a patient, as this may cause patient
injury or death.
29. WARNING: Unplug the AC power cord before opening the pump housing to avoid risk of electric shock,
which could result in death or serious injury.
30. WARNING: Potentially dangerous voltages are exposed when the pump housing is open and the AC power
cord is plugged in. These voltages are present in the transformer and AC fuse on the CPU/Power supply
board and the RFI line filter where the power cord connects. To avoid electric shock with potential for
severe injury or death, do not touch these areas when the power cord is plugged in.
Summary of Contents for 3000/500
Page 1: ...0473 Technical Service Manual Model 3000 500 and Micro 3100 505 Volumetric Infusion Pump...
Page 2: ......
Page 14: ...I n t r o d u c t i o n Chapter 1...
Page 18: ...S p e c i f i c a t i o n Chapter 2...
Page 29: ...O v e r v i e w o f P u m p O p e r a t i o n I n i t i a l C h e c k s Chapter 3...
Page 47: ...V o l u m e t r i c P u m p M e n u s Chapter 4...
Page 79: ...M e c h a n i c a l S y s t e m s Chapter 5...
Page 95: ...E l e c t r o n i c C i r c u i t s Chapter 6...
Page 96: ...M a i n t e n a n c e P r o c e d u r e s T r o u b l e s h o o t i n g Chapter 7...
Page 151: ...I l l u s t r a t e d P a r t s L i s t Chapter 9...
Page 178: ...E l e c t r i c a l D i a g r a m s Chapter 10...
Page 189: ...W a v e f o r m D i a g r a m s Chapter 11...
Page 198: ...R e m o t e M o n i t o r i n g C o n t r o l o f t h e p u m p Chapter 12...
Page 225: ...I n d e x...