![Graseby 3000/500 Manual Download Page 60](http://html1.mh-extra.com/html/graseby/3000-500/3000-500_manual_2257430060.webp)
4 — 1 3
G r a s e b y M e d i c a l L t d .
4
3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l
I s s u e A ( A p r i l 2 0 0 2 )
V o l u m e t r i c I n f u s i o n P u m p s
Snapshot screen-a - system state codes
This list shows the possible system/state codes shown on the top right
hand line of a Snapshot screen-a:
00
No action, initialization value.
01
Power Fail (Battery) Detected.
02
Motor Monitor Communication time-out error.
03
Motor Monitor RAM error.
09
Motor Monitor Watchdog error.
10
CPU1 ROM Checksum error.
11
Motor Monitor Checksum error.
12
CPU1 RAM failure.
13
*
NV RAM checksum mismatch.
14
Main audio alarm Failure.
15
Motor fast error detected by Motor Monitor.
16
Motor slow error detected by Motor Monitor.
17
Motor disable test failure.
19
Air detect test failure.
20
No flow above pump circuit error - return pump to SIMS
Graseby or Service centre.
21
Display board communications error.
22
Display digit error, 0-9.
23
Display LED error, 0-6.
24
Display Board ROM error.
25
Display Board RAM error.
28
Motor stall/interrupter.
30
Warm Boot.
31
Micro/Macro discrepancy error.
32
Pressure Plate Position Switch change during RUN.
33
*
Corrupted Triple Redundant variable in NV RAM.
34
LCD Display Failure.
35
Undefined Interrupt occurred.
36
Display field overflow error.
37
Display field parameter error.
38
Nonvolatile RAM battery low.
39
Stack overflow.
40
Pressure Plate Position Switch =1 when door is open.
41
Pressure >1 psi when door is open.
42
Safety clip in place when door is open.
This is the system
state code
Code
Meaning
list continues ...
* see
Faulty Component Identification
in the Troubleshooting section on page 7-17
Summary of Contents for 3000/500
Page 1: ...0473 Technical Service Manual Model 3000 500 and Micro 3100 505 Volumetric Infusion Pump...
Page 2: ......
Page 14: ...I n t r o d u c t i o n Chapter 1...
Page 18: ...S p e c i f i c a t i o n Chapter 2...
Page 29: ...O v e r v i e w o f P u m p O p e r a t i o n I n i t i a l C h e c k s Chapter 3...
Page 47: ...V o l u m e t r i c P u m p M e n u s Chapter 4...
Page 79: ...M e c h a n i c a l S y s t e m s Chapter 5...
Page 95: ...E l e c t r o n i c C i r c u i t s Chapter 6...
Page 96: ...M a i n t e n a n c e P r o c e d u r e s T r o u b l e s h o o t i n g Chapter 7...
Page 151: ...I l l u s t r a t e d P a r t s L i s t Chapter 9...
Page 178: ...E l e c t r i c a l D i a g r a m s Chapter 10...
Page 189: ...W a v e f o r m D i a g r a m s Chapter 11...
Page 198: ...R e m o t e M o n i t o r i n g C o n t r o l o f t h e p u m p Chapter 12...
Page 225: ...I n d e x...