![Graseby 3000/500 Manual Download Page 210](http://html1.mh-extra.com/html/graseby/3000-500/3000-500_manual_2257430210.webp)
Graseby Medical L t d
Volumetric Infusion Pumps
12 - 12
Issue 1(April 2002)
3000/3100 and 500/505 Service Manual
The pump will also respond to the following “CI” or interface/control
commands:
CI (Interface)
Command Code
Name
Parameter Needed
B
Silence Key Keypress
C
Infused Volume Clear
E
Time and Date Set
ü
F
Pump On or Off.
ü
K
Keypad Lockout Setup
ü
L
VTBI Set
ü
N
Open Top Valve
ü
O
Run Mode Enter (Begin pumping).
P
Occlusion Pressure Set (2,5, or 10 psi).
ü
R
Rate Value Set
ü
S
Hold Mode Enter (Stop pumping).
T
Titrate (Rate change while pumping).
ü
X
User LCD Text Setup
ü
Command Parameters
Any characters following the command code and preceding the CRC
characters are parameters specific to the command code. There may be
a minimum of zero and a maximum of 17 parameter characters.
Responses from the pump are described in Table 1 and the section
entitled 'Return Record Format'. When the pump receives a CI
command that requires a parameter, and that parameter is out of
range or missing, the pump will indicate ‘invalid parameter’ in the
‘return code field.’
CRC
Immediately following the parameter area (if any), a four character
CRC is given. The CRC is calculated as a 16-bit value and includes the
record header through the parameter characters. The record
terminator, ">", and CRC characters are not included in the CRC.
When the pump receives a message where the CRC field does not
match calculated message CRC, the pump will indicate ‘Invalid CRC
Field’ in the ‘return code field.’
The CRC table and algorithm used are shown below in a 'C' code
example:
unsigned int crctable[256] =
{
0x0000, 0x1189, 0x2312, 0x329B, 0x4624, 0x57AD, 0x6536, 0x74BF,
0x8C48, 0x9DC1, 0xAF5A, 0xBED3, 0xCA6C, 0xDBE5, 0xE97E, 0xF8F7,
0x0919, 0x1890, 0x2A0B, 0x3B82, 0x4F3D, 0x5EB4, 0x6C2F, 0x7DA6,
0x8551, 0x94D8, 0xA643, 0xB7CA, 0xC375, 0xD2FC, 0xE067, 0xF1EE,
0x1232, 0x03BB, 0x3120, 0x20A9, 0x5416, 0x459F, 0x7704, 0x668D,
Summary of Contents for 3000/500
Page 1: ...0473 Technical Service Manual Model 3000 500 and Micro 3100 505 Volumetric Infusion Pump...
Page 2: ......
Page 14: ...I n t r o d u c t i o n Chapter 1...
Page 18: ...S p e c i f i c a t i o n Chapter 2...
Page 29: ...O v e r v i e w o f P u m p O p e r a t i o n I n i t i a l C h e c k s Chapter 3...
Page 47: ...V o l u m e t r i c P u m p M e n u s Chapter 4...
Page 79: ...M e c h a n i c a l S y s t e m s Chapter 5...
Page 95: ...E l e c t r o n i c C i r c u i t s Chapter 6...
Page 96: ...M a i n t e n a n c e P r o c e d u r e s T r o u b l e s h o o t i n g Chapter 7...
Page 151: ...I l l u s t r a t e d P a r t s L i s t Chapter 9...
Page 178: ...E l e c t r i c a l D i a g r a m s Chapter 10...
Page 189: ...W a v e f o r m D i a g r a m s Chapter 11...
Page 198: ...R e m o t e M o n i t o r i n g C o n t r o l o f t h e p u m p Chapter 12...
Page 225: ...I n d e x...