7 — 1 5
G r a s e b y M e d i c a l L t d .
3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l
I s s u e A ( A p r i l 2 0 0 2 )
V o l u m e t r i c I n f u s i o n P u m p s
7
After the pump has delivered 49 mL (25.0 mL on
3100/505) the
non-
insistent
audio alarm sounds.
Stop the stopwatch.
Verify that the following display appears in the Message Display:
Confirm that the volume to be infused display is 0.
Confirm that the rate changes to 3 mL/h, or the designated KVO
setting in a pump with V0.71 software. Press
Hold
.
Check the burette and verify a volume of 49 mL ± 1 mL on 3000/500. If
checking a 3100/505, verify that the burette holds 25 mL ± 0.5 mL.
Remember to take the measurement from the bottom of the meniscus.
Disconnect and drain the burette.
Rate Accuracy:
Check the stopwatch and verify that the alarm occurred at the
correct point:
Pump model
Alarm should occur after...
3000/500
2 minutes 56.6 seconds ± 3 seconds.
3100/505
15 minutes ± 18 seconds.
15. Testing the battery
Graseby Medical recommends that the condition of the lead
acid battery is checked at least annually
Method 1
When the pump is powered on, the battery is always tested. In
addition, the
Options
key provides access to the Battery Test option
so that the approximate battery voltage can be tested without
powering down the pump.
To run the Battery Test:
1. With the pump powered up, disconnect from the AC mains supply
and press
Hold
.
2. Press the
Options
key then
4
, or press
Options
until the following
message is displayed:
3. Press the key to start the test. The pump turns on all LEDs, and
the rate and volume to be infused displays all zeros to form a
known battery load.
Summary of Contents for 3000/500
Page 1: ...0473 Technical Service Manual Model 3000 500 and Micro 3100 505 Volumetric Infusion Pump...
Page 2: ......
Page 14: ...I n t r o d u c t i o n Chapter 1...
Page 18: ...S p e c i f i c a t i o n Chapter 2...
Page 29: ...O v e r v i e w o f P u m p O p e r a t i o n I n i t i a l C h e c k s Chapter 3...
Page 47: ...V o l u m e t r i c P u m p M e n u s Chapter 4...
Page 79: ...M e c h a n i c a l S y s t e m s Chapter 5...
Page 95: ...E l e c t r o n i c C i r c u i t s Chapter 6...
Page 96: ...M a i n t e n a n c e P r o c e d u r e s T r o u b l e s h o o t i n g Chapter 7...
Page 151: ...I l l u s t r a t e d P a r t s L i s t Chapter 9...
Page 178: ...E l e c t r i c a l D i a g r a m s Chapter 10...
Page 189: ...W a v e f o r m D i a g r a m s Chapter 11...
Page 198: ...R e m o t e M o n i t o r i n g C o n t r o l o f t h e p u m p Chapter 12...
Page 225: ...I n d e x...