Graseby Medical Ltd.
Volumetric Infusion Pumps
12 - 26
Issue 1(November 2001)
3000/3100 and 500/505 Serv ice Manual
Example Commands
Version 0.68 and above
Command
Description:
Command Sent by
Host:
Pump Response:
Data query
<09Dxxxx>
<20
RAAAAAA
aaaa.abbbb.bccc.c
XXXX
>
Enable CI commands
with 60 s timeout
<0BI60xxxx>
<20
RAAAAAA
aaaa.abbbb.bccc.c
XXXX
>
Set Limit to 1000 ml
<0DL1000xxxx>
<20
RAAAAAA
aaaa.a1000.0ccc.c
XXXX
>
Run
<09Oxxxx>
<20
RAAAAAA
aaaa.abbbb.bccc.c
XXXX
>
Set Rate to 100 ml/h
<0CR100xxxx>
<20
RAAAAAA
aaaa.abbbb.b100.0
XXXX
>
Hold
<09Sxxxx>
<20
RAAAAAA
aaaa.abbbb.bccc.c
XXXX
>
where:
R
= return code
XXXX
= CRC
AAAAAA
= alarm bytes
aaaa.a
= volume infused
bbbb.b
= volume limit
ccc.c
= rate in mL/hr
Summary of Contents for 3000/500
Page 1: ...0473 Technical Service Manual Model 3000 500 and Micro 3100 505 Volumetric Infusion Pump...
Page 2: ......
Page 14: ...I n t r o d u c t i o n Chapter 1...
Page 18: ...S p e c i f i c a t i o n Chapter 2...
Page 29: ...O v e r v i e w o f P u m p O p e r a t i o n I n i t i a l C h e c k s Chapter 3...
Page 47: ...V o l u m e t r i c P u m p M e n u s Chapter 4...
Page 79: ...M e c h a n i c a l S y s t e m s Chapter 5...
Page 95: ...E l e c t r o n i c C i r c u i t s Chapter 6...
Page 96: ...M a i n t e n a n c e P r o c e d u r e s T r o u b l e s h o o t i n g Chapter 7...
Page 151: ...I l l u s t r a t e d P a r t s L i s t Chapter 9...
Page 178: ...E l e c t r i c a l D i a g r a m s Chapter 10...
Page 189: ...W a v e f o r m D i a g r a m s Chapter 11...
Page 198: ...R e m o t e M o n i t o r i n g C o n t r o l o f t h e p u m p Chapter 12...
Page 225: ...I n d e x...