![Graseby 3000/500 Manual Download Page 113](http://html1.mh-extra.com/html/graseby/3000-500/3000-500_manual_2257430113.webp)
7 — 1 7
G r a s e b y M e d i c a l L t d .
3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l
I s s u e A ( A p r i l 2 0 0 2 )
V o l u m e t r i c I n f u s i o n P u m p s
7
5. Using the Volumetric Event History Download Software
confirm that the times up to the first Low Battery Plug In Cord
alarm, and time difference between this alarm and the Battery
Too Low Plug In Cord alarm, is within the following
specification.
Pass:
A battery shall be deemed to have passed its test if the unit runs for a
minimum of 51/2 hours on DC power, at a rate of 100ml/hr (99ml/hr
for 505/3100) before the first Low Battery Alarm Plug In Cord is
triggered.
There must be a minimum of 30 minutes between this alarm and the
Dead Battery Plug In Cord alarm that will occur before the pump
powers down.
In use this 30 minutes would allow up to 5 sets of Low Battery Plug
In Cord alarms to be raised. When cancelled the Low Battery alarm
will be triggered again every 6 minutes if the AC power supply has
not been connected.
Fail:
A faulty battery would be one that does not meet this specification. In
this case the battery should be replaced.
It is recommended that pumps should be run overnight as this would
avoid creating a disturbance to staff caused by the Low Battery Plug
In Cord and Dead Battery Plug In Cord alarms that occur before the
units power down.
Event history should then be downloaded and analysed to confirm
that the battery time periods meet the requirements above.
Summary of Contents for 3000/500
Page 1: ...0473 Technical Service Manual Model 3000 500 and Micro 3100 505 Volumetric Infusion Pump...
Page 2: ......
Page 14: ...I n t r o d u c t i o n Chapter 1...
Page 18: ...S p e c i f i c a t i o n Chapter 2...
Page 29: ...O v e r v i e w o f P u m p O p e r a t i o n I n i t i a l C h e c k s Chapter 3...
Page 47: ...V o l u m e t r i c P u m p M e n u s Chapter 4...
Page 79: ...M e c h a n i c a l S y s t e m s Chapter 5...
Page 95: ...E l e c t r o n i c C i r c u i t s Chapter 6...
Page 96: ...M a i n t e n a n c e P r o c e d u r e s T r o u b l e s h o o t i n g Chapter 7...
Page 151: ...I l l u s t r a t e d P a r t s L i s t Chapter 9...
Page 178: ...E l e c t r i c a l D i a g r a m s Chapter 10...
Page 189: ...W a v e f o r m D i a g r a m s Chapter 11...
Page 198: ...R e m o t e M o n i t o r i n g C o n t r o l o f t h e p u m p Chapter 12...
Page 225: ...I n d e x...