4 — 5
G r a s e b y M e d i c a l L t d .
4
3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l
I s s u e A ( A p r i l 2 0 0 2 )
V o l u m e t r i c I n f u s i o n P u m p s
KVO rate entry
If this is enabled, then the user can set their own KVO rate. If it is
disabled, the default KVO rate (as shown in the specifications) is used.
Occlusion default setting
Use this option to set the occlusion default setting that should be
applied when the pump is switched on. Press * to change the default.
The user can press the
Occlusion Alarm Setting
key to change the
occlusion level for a particular infusion regardless of this default
parameter setting. However the default setting is reapplied when the
pump is switched off and on again.
Secondary stop
With this parameter enabled, the pump stops when a Secondary
infusion is completed. It sounds an insistent alarm and displays a
message to say that the secondary infusion is complete.
If Secondary stop is disabled, when a secondary infusion is complete,
the pump sounds a non-insistent alarm twice and then automatically
restarts the infusion at the primary rate.
This parameter is independent of the setting of the Bolus parameter
and only affects a Secondary infusion; at the end of a Bolus infusion
the pump always reverts to the previously programmed Primary rate.
Bolus
When this parameter is enabled, the user is given a choice of whether
a Secondary or a Bolus infusion is to be run when they press the
Secondary/Bolus
key. If Bolus is chosen, the volume infused is
deducted from the Primary VTBI. If Secondary is chosen, the
Secondary volume infused is kept as a separate total and does not
affect the Primary VTBI.
If Bolus is disabled, the user is not given a choice when they press the
Secondary/Bolus
key: they can only make settings for a Secondary
infusion.
Quick Rate Change
In software versions before 0.71, the Quick Rate Change parameter
was always enabled, and could not be disabled. From version 0.71, the
parameter may be enabled or disabled on the Technician Menu.
Enabling this feature allows users to use the
Options
key to put the
pump into a mode in which they can change the rate quickly while the
pump is running or on hold.
Rate Taper
Enabling this parameter allows the user to taper the infusion rate,
either up or down, for the administration of Total Parenteral Nutrition
(TPN), Total Parenteral Admixture (Three-in-One) or other applicable
therapies.
Summary of Contents for 3000/500
Page 1: ...0473 Technical Service Manual Model 3000 500 and Micro 3100 505 Volumetric Infusion Pump...
Page 2: ......
Page 14: ...I n t r o d u c t i o n Chapter 1...
Page 18: ...S p e c i f i c a t i o n Chapter 2...
Page 29: ...O v e r v i e w o f P u m p O p e r a t i o n I n i t i a l C h e c k s Chapter 3...
Page 47: ...V o l u m e t r i c P u m p M e n u s Chapter 4...
Page 79: ...M e c h a n i c a l S y s t e m s Chapter 5...
Page 95: ...E l e c t r o n i c C i r c u i t s Chapter 6...
Page 96: ...M a i n t e n a n c e P r o c e d u r e s T r o u b l e s h o o t i n g Chapter 7...
Page 151: ...I l l u s t r a t e d P a r t s L i s t Chapter 9...
Page 178: ...E l e c t r i c a l D i a g r a m s Chapter 10...
Page 189: ...W a v e f o r m D i a g r a m s Chapter 11...
Page 198: ...R e m o t e M o n i t o r i n g C o n t r o l o f t h e p u m p Chapter 12...
Page 225: ...I n d e x...