Adding more than one Powerbar 5.1 or Doser 2.1?
When connecting multiple PB5.1 or D2.1 devices, the numbering for each
device must be assigned. To configure the numbering of these devices,
select the PAB-device, click on CONFIGURATION, and set the Powerbar
socket or dosing pump numbering. You can see how this is done by clicking
the how-to video link above.
TO DO:
Probe Calibration
In order for probes to remain accurate, they must undergo calibration on a regular basis. New
probes must always be calibrated before being put to use. The steps for calibrating each
probe will vary. There are however some probes which do not require calibration. The
temperature probe for example, is the only probe from the P4 set which does not require any
calibration.
NOTE
Probe calibration can be done through the ProfiLux 4 display. No need to use
GHL Control Center for this process.
Best calibration practices
In order to acheive the most accurate probe readings, proper calibration
must be performed. Here are some tips to go by:
ALWAYS calibrate new probes; pH, redox, Conductivity.
Calibrate every 1-3 months
Rinse and dry the probe to remove any traces of tank water BEFORE
starting calibration; failiure to do so could result in cross
contamination of calibration fluids.
Make sure probe and eye is clean and free from any detritus or
foreign material.
Wipe probe clean BEFORE dipping into another calibration fluid; this
will avoid cross contamination.
Make sure NO microbubbles get stuck underneath the probe.