fixed temp. Value. For more info on Conductivity and temperature
compensation, please see page 54 of the P4 Programming Guide.
NOTE: Listen for the 3-beeps
During calibration, the P4 will give the probe 900s (15 minutes) for the ADC
values to stabilize. For steps 8 and 9, when you hear 3 beeps, the calibration
step is complete. The amount of time it will take to hear these beeps will
depend on probe age and set Calibration Tolerance.
1.
Press any of the arrow keys on
the P4 and select SENSOR
SETTINGS.
2.
Select CONDUCT.1 or desired
conductivity probe to calibrate.
3.
Select CALIBRATION.
4.
Select (1) for CALI. TOLERANCE
if probe is brand new or less
than 1 year old.
5.
Select 50mS for CALIBRA.VALUE 2.
6.
Select if the calibration process will use the temperature compensation readings from a
temperature probe or a manually input temperature value.
7.
Enter the exact temperature of
the calibration fluid. If bottle is
still floating in the aquarium, this
temperature will be the same temperature as your tank.
8.
Take the probe out of the tank and let it sit out in the air,
then press the (√) key to begin the countdown. When you
hear 3 beeps, proceed to the next step.
9.
Dip the probe in 50mS calibration fluid and press the (√) key. When you hear 3 beeps,
select YES to save calibration.
10.
Wipe the probe and put it back into the aquarium.