September 2007
6-42
General Programming
TEST RECORDER Menu
The
Test Recorder?
option lets you check the accuracy of the XMO2
4-20 mA analog output. To use this option, refer to the menu map in
Figure C-6 on page C-6, and proceed as follows:
Note:
Before testing the 4-20 mA analog output, be sure the output is
properly scaled (see pages 5-10 to 5-11, or see page 6-36) and
calibrated (see page 7-32).
The next prompt lists whichever choice was selected the last time the
menu was accessed:
•
Enter mA
- tests the 4-20 mA output to a specified mA value
•
Enter Units
- tests the 4-20 mA output to a specified % oxygen
Based on the selection made above, go to the appropriate section.
ENTER mA Option
If you selected the
Enter mA?
option, enter the desired analog output
in mA at the next prompt.
The 4-20 mA analog output is forced to the value specified above.
Verify that your analog output device is displaying the correct value.
This completes the prompts for the
Enter mA?
option, and the XMO2
returns to the main
General
menu.
GENERAL MENU
Press
[Y]
or
[Enter]
to enter the
Test Recorder?
menu, or press
[N]
to move to the next
General
Menu
option.
Test Recorder?
Select Value Type:
Press
[N]
until the desired option
is displayed, then press
[Y]
or
[Enter]
to proceed.
(Enter mA?) or (Enter Units?)
Enter mA Output mA:
To accept the current value, press
[Y]
or
[Enter]
. To change the
current value, use the numeric
keys to enter a new value and
press
[Y]
or
[Enter]
twice.
mA [xx.xx]:
mA Output @ xx.xx mA
When ready, press
[Y]
or
[Enter]
to return to the
General
menu:
Press [YES]
Summary of Contents for XMO2
Page 8: ...Chapter 1 ...
Page 18: ...Chapter 2 ...
Page 30: ...Chapter 3 ...
Page 41: ...Chapter 4 ...
Page 52: ...Chapter 5 ...
Page 66: ...Chapter 6 ...
Page 114: ...Chapter 7 ...
Page 146: ...Chapter 8 ...
Page 159: ...Appendix A ...
Page 165: ...Appendix B ...
Page 177: ...Appendix C ...