General Programming
6-33
September 2007
DriftCal Enabled and Two
Gases Selected (cont.)
The
Reset DriftCal
option:
This option is used to correct erratic readings from a faulty calibration
by deleting all calibration data stored from previous calibrations.
Caution!
This option should only be used to clear a faulty calibration
that resulted in the
“DriftCal Error - Check Cal Gases”
screen error message.
At the next prompt, selecting
Yes
will perform the reset, while
selecting
No
will abort the process.
Caution!
Be absolutely sure that you want to erase all of the stored
Drift Curve
data before you select and confirm a
Yes
response below.
This completes the prompts for the
Reset DriftCal
option, and the
XMO2 returns to the
Two-Gas DriftCal
menu.
DRIFTCAL MENU
Press
[Y]
or
[Enter]
to continue, or
press
[N]
to proceed to the next
menu option.
Reset DriftCal?
WARNING: Reset DriftCal?
Press
[N]
to select the desired
response. Then, press
[Y]
or
[Enter]
confirm your choice.
[YES] no