September 2007
6-8
General Programming
BACKGROUND GRID
Option
Prior to shipment, the XMO2 factory calibration data is programmed
into the
Background Grid
menu. If the XMO2 displays erroneous
readings, verify that the calibration data in the
Background Grid
menu matches that on your calibration data sheet (see page 8-6 for a
sample sheet). If the data does not match, contact the factory.
IMPORTANT:
Always consult the factory before proceeding with
this option.
The
Background Grid
option is used to enter calibration data for an
XMO2 that is configured with background gas compensation. Data is
entered as a series of calibration curves, each of which represents a
particular background gas (the data for your curves is supplied on an
XMO2 calibration sheet similar to the one shown on page 8-6). To
use the
Background Grid
option, refer to the menu map in Figure C-2
on page C-2 and complete the following steps:
Note:
The
Background Grid
option appears only if
Pressure
Compensation
is
OFF
and
Background Compensation is
ON
,
as specified in the
System Mode
menu (see page 6-4).
At the
# Curves
prompt, specify the number of calibration data curves
to be entered. Each curve represents a specific concentration of a
particular background gas.
IMPORTANT:
The
# Curves
for your unit is listed on its calibration
data sheet (see page 8-6 for a sample sheet), and the
value should not be changed.
The next prompt appears only if the
Display Background Gas
option
is enabled in the
Set Background Display
branch of the Advanced
Menu (see Chapter 7,
Advanced Programming
). The
Curve #n
prompt allows you to enter a fixed background gas concentration for
one of the curves in the background grid.
Note:
The
gas type
displayed in the
Curve #n
prompt is specified in
the
Select Gas Type
option of the
Advanced Menu
. The
choices are %Bkgd (generic),
%H2
,
%N2
(used as an example
in this manual),
%SO2
, and
%CO2
. Refer to Chapter 7,
Advanced Programming
, for more details.
CALIBRATE SYSTEM
Press
[Y]
or
[Enter]
to enter the
Background Grid
menu.
Background Grid?
CALIBRATE GRID
Accept the current value by
pressing
[Y]
or
[Enter]
.
# Curves [x]:
Summary of Contents for XMO2
Page 8: ...Chapter 1 ...
Page 18: ...Chapter 2 ...
Page 30: ...Chapter 3 ...
Page 41: ...Chapter 4 ...
Page 52: ...Chapter 5 ...
Page 66: ...Chapter 6 ...
Page 114: ...Chapter 7 ...
Page 146: ...Chapter 8 ...
Page 159: ...Appendix A ...
Page 165: ...Appendix B ...
Page 177: ...Appendix C ...