Basic Programming
5-7
September 2007
QUICK ZERO Option
The
Quick Zero
option
uses a zero gas of known concentration to
perform a quick field recalibration the XMO2. The XMO2 compares
the data from the new field calibration to the original factory
calibration data, and stores the difference in a
Drift Curve
.
Note:
This option appears only if
DriftCal
is
ON
and a
two-gas
calibration
has been selected.
To use the
Quick Zero
option, complete the following steps:
1.
Using the sample system controls, stop the flow of sample gas to
the XMO2 inlet port and initiate a flow of the same zero gas
specified on the XMO2 Calibration Sheet. Establish the same flow
rate and pressure conditions used for the sample gas, and allow the
zero gas to flow through the XMO2 for at least three minutes.
2.
Perform the zero gas calibration, press
[N]
until the following
prompt appears. Then, proceed as follows:
You may now do one of the following:
•
Press
[N]
as many times as necessary to select another
Basic Menu
option. Then, press
[Y]
and proceed to the appropriate section for
instructions.
•
Press
[N]
repeatedly until the
Resume?
option appears. Then, press
[Y]
to return to
Operate Mode
.
BASIC MENU
Press
[Y]
or
[Enter]
to initiate the
zero gas calibration.
Quick Zero?
BASIC MENU
This screen appears while the
calibration is taking place.
DriftCal In Progress...
BASIC MENU
The zero gas calibration has been
completed.
Quick Zero?
Summary of Contents for XMO2
Page 8: ...Chapter 1 ...
Page 18: ...Chapter 2 ...
Page 30: ...Chapter 3 ...
Page 41: ...Chapter 4 ...
Page 52: ...Chapter 5 ...
Page 66: ...Chapter 6 ...
Page 114: ...Chapter 7 ...
Page 146: ...Chapter 8 ...
Page 159: ...Appendix A ...
Page 165: ...Appendix B ...
Page 177: ...Appendix C ...