General Programming
DriftCal Enabled and One
Gas Selected (cont.)
The
Perform Offset
option:
This option performs an one-gas calibration of the XMO2 using the
offset gas and time interval specified in the
Set Offset Gas
option.
IMPORTANT:
Be sure the XMO2 is connected to the correct offset
gas at the correct flow rate prior to using the
Perform Offset
option. Refer to Chapter 4,
Field
Calibration
, for details.
If
[Y]
was pressed above, the XMO2 begins taking offset gas readings.
Approximately every 2.5 sec, it displays the % oxygen and the time
remaining for the calibration.
After the specified calibration time has expired, the XMO2 records
the calibration data and displays the new calibration.
Note:
If the XMO2 does not reach equilibrium before the calibration
time expires, the calibration is rejected and the error message,
“Offset Gas Calibration Error-Check Cal Gases”
is displayed.
To correct the error, erase the DriftCal data in the
Reset
DriftCal
menu (see page 6-33).
This completes the prompts for the
Perform Offset
option, and the
XMO2 returns to the
One-Gas DriftCal
menu.
DRIFTCAL MENU
Press
[Y]
or
[Enter]
to perform an
offset calibration, or press
[N]
to
proceed to the next menu option.
Perform Offset?
Cal Gas ON (M:SS)
xx.xx %O2
DriftCal in Progress
Measuring Drift x.xx %O2
Summary of Contents for XMO2
Page 8: ...Chapter 1 ...
Page 18: ...Chapter 2 ...
Page 30: ...Chapter 3 ...
Page 41: ...Chapter 4 ...
Page 52: ...Chapter 5 ...
Page 66: ...Chapter 6 ...
Page 114: ...Chapter 7 ...
Page 146: ...Chapter 8 ...
Page 159: ...Appendix A ...
Page 165: ...Appendix B ...
Page 177: ...Appendix C ...