September 2007
6-16
General Programming
PRESSURE GRID Option
(cont.)
•
Numeric Entry/
[Y]
- enter tabulated mV value
•
Numeric Entry/
[N]
- XMO2 interpolates mV value
•
Introduce Gases/
[Y]
- use appropriate calibration gas
•
Introduce Gases/
[N]
- XMO2 interpolates mV value
If you selected
Numeric Entry
and
[Y]
:
If you selected
Numeric Entry
at the
Calibrate Mode
prompt and
[Y]
at the
Enter Point?
prompt, the
O2 mV
prompt lets you enter the
corresponding O2 mV value for the current %O2 and pressure values.
After responding to the above prompt, you are returned to the
Enter
Point?
prompt on page 6-15.
If you selected
Numeric Entry
and
[N]
:
If you selected
Numeric Entry
at the
Calibrate Mode
prompt and
[N]
at the
Enter Point?
prompt, the
Interpolate Point?
prompt is used to
have the XMO2 perform a linear interpolation to determine the
corresponding mV value for the current %O2 and pressure values.
Note:
The XMO2 linear interpolation capability is useful when a
data curve is missing one or more of its data points.
After the XMO2 performs the linear interpolation, you are returned to
the
Enter Point?
prompt on page 6-15.
x.xx %O2, xxx.x mmHg
To accept the current value, press
[Y]
or
[Enter]
. To change the
current value, use the numeric
keys to enter a new value and
press
[Y]
or
[Enter]
twice.
O2 mV [-x.x]:
x.xx %O2, xxx.x mmHg
Press
[Y]
or
[Enter]
to begin the
interpolation, or press
[N]
to
return to the
Enter Point?
prompt.
Interpolate Point?
Summary of Contents for XMO2
Page 8: ...Chapter 1 ...
Page 18: ...Chapter 2 ...
Page 30: ...Chapter 3 ...
Page 41: ...Chapter 4 ...
Page 52: ...Chapter 5 ...
Page 66: ...Chapter 6 ...
Page 114: ...Chapter 7 ...
Page 146: ...Chapter 8 ...
Page 159: ...Appendix A ...
Page 165: ...Appendix B ...
Page 177: ...Appendix C ...