Startup & Operation
3-9
September 2007
Changing the 4-20 mA
Analog Output Range
(cont.)
6.
At the last screen on page 3-8, do one of the following:
•
Press
[Y]
or
[Enter]
to accept the current high input value
(100.00 in the above example). Then, go to Step 8.
•
Use the numeric keys to enter a new high input value (25.00 in
the example that follows). Then, go to Step 7.
7.
If you entered a new high input value:
8.
Complete the procedure as follows:
The range of the 4-20 mA analog output has now been changed. Be
sure to revise your XMO2
Calibration Sheet
to show the new range.
For additional information, or if programming assistance is required,
refer to Chapter 5,
Basic Programming
.
mA Output 4 mA Value
Press
[Y]
or
[Enter]
to accept the
new high (20 mA) input value.
%O2 [100.00]:25.00
BASIC MENU
Press
[N]
repeatedly until the
following menu option appears:
Set High Input Value?
BASIC MENU
Press
[Y]
or
[Enter]
to return to
Operate Mode
.
Resume?