6-38
General Programming
COMP INPUT Option
The
Comp Input
option lets you monitor the background gas
compensation input signal. The acceptable range for this signal is
0 mV to +500 mV.
The current background gas
Comp Input
signal is displayed (if the
displayed value is very different from that on the Calibration Sheet,
contact the factory for assistance).
This completes the prompts for the
Comp Input
option, and the
XMO2 returns to the
Test Inputs?
menu.
PRESSURE INPUT Option
The
Pressure Input
option lets you monitor the pressure sensor input
compensation signal. The acceptable range for this signal is -512 mV
to +512 mV.
The current
Pressure Input
signal from the optional pressure sensor is
displayed.
This completes the prompts for the
Pressure Input
option, and the
XMO2 returns to the
Test Inputs?
menu.
TEST INPUTS
Press
[Y]
or
[Enter]
to view the
Comp input, or press
[N]
to move
to the next
Test Inputs?
option.
Comp Input?
Comp Input
When ready, press any key to
return to the
Test Inputs?
menu:
xxx.x Bkgd mV
TEST INPUTS
Press
[Y]
or
[Enter]
to view the
Pressure input, or press
[N]
to
move to the next
Test Inputs?
option.
Pressure Input?
Pressure Input
When ready, press any key to
return to the
Test Inputs?
menu:
-xx.x mmHg mV
Summary of Contents for XMO2
Page 8: ...Chapter 1 ...
Page 18: ...Chapter 2 ...
Page 30: ...Chapter 3 ...
Page 41: ...Chapter 4 ...
Page 52: ...Chapter 5 ...
Page 66: ...Chapter 6 ...
Page 114: ...Chapter 7 ...
Page 146: ...Chapter 8 ...
Page 159: ...Appendix A ...
Page 165: ...Appendix B ...
Page 177: ...Appendix C ...