Basic Programming
5-11
September 2007
SET HIGH INPUT VALUE
Option
The
Set High Input Value option
lets you specify the %O
2
value in the
gas that corresponds to a 20.00 mA current at the analog output.
Note:
This option appears if
DriftCal
is
ON
or
OFF
and if a
one- or
two-gas calibration
has been selected.
To use the
Set High Input Value
option, proceed as follows:
1.
Press
[N]
until the following prompt appears:
2.
At the above screen, do one of the following:
•
Press
[Y]
or
[ENTER]
to accept the current high input value
(100.00 in the above example). Then, go to Step 4.
•
Use the numeric keys to enter a new high input value (25.00 in
the example that follows). Then, go to Step 3.
3.
If you entered a new high input value:
4.
Complete the procedure as follows:
You may now do one of the following:
•
Press
[N]
as many times as necessary to select another
Basic Menu
option. Then, press
[Y]
and proceed to the appropriate section for
instructions.
•
Press
[N]
repeatedly until the
Resume?
option appears. Then, press
[Y]
to return to
Operate Mode
.
BASIC MENU
Press
[Y]
or
[ENTER]
to set the
high (20 mA) input value.
Set High Input Value?
mA Output 20 mA Value
Go to Step 2.
%O2 [100.00]:
mA Output 20 mA Value
Press
[Y]
or
[ENTER]
to accept the
new high (20 mA) input value.
%O2 [100.00]:25.00
BASIC MENU
The
High Input Value
has been
set.
Set High Input Value?
Summary of Contents for XMO2
Page 8: ...Chapter 1 ...
Page 18: ...Chapter 2 ...
Page 30: ...Chapter 3 ...
Page 41: ...Chapter 4 ...
Page 52: ...Chapter 5 ...
Page 66: ...Chapter 6 ...
Page 114: ...Chapter 7 ...
Page 146: ...Chapter 8 ...
Page 159: ...Appendix A ...
Page 165: ...Appendix B ...
Page 177: ...Appendix C ...